Breast Transillumination, Electrical Impedance Scanning (EIS), and Elastography

Number: 0386

Table Of Contents

Policy
Applicable CPT / HCPCS / ICD-10 Codes
Background
References


Policy

Scope of Policy

This Clinical Policy Bulletin addresses breast transillumination, electrical impedance scanning (EIS), and elastography.

  1. Experimental and Investigational

    The following procedures are considered experimental and investigational because the effectiveness of these approaches has not been established:

    1. Breast elastography by any method (i.e., ultrasound or magnetic resonance) 
    2. Combined shear-wave elastography with ultrasonography for detection of breast cancer
    3. Electrical impedance scanning (EIS) of the breast for distinguishing benign from malignant breast lesions
    4. Non-elastographic quantitative ultrasound tissue characterization for screening and diagnosis of breast cancer and for all other indications 
    5. Optical coherence micro-elastography for identifying malignant tissue and lymph node involvement during breast-conserving surgery
    6. Opto-acoustic imaging of relative blood oxygen saturation and total hemoglobin for diagnosis of breast cancer
    7. Transillumination (optical imaging, light scanning or diaphanography) of the breast for detection of breast cancer.
  2. Related Policies


Table:

CPT Codes / HCPCS Codes / ICD-10 Codes

Code Code Description

Information in the [brackets] below has been added for clarification purposes.   Codes requiring a 7th character are represented by "+":

CPT codes not covered for indications listed in the CPB:

Breast transillumination -no specific code
0351T - 0352T Optical coherence tomography of breast or axillary lymph node, excised tissue, each specimen
0353T - 0354T Optical coherence tomography of breast, surgical cavity
0689T Quantitative ultrasound tissue characterization (non-elastographic), including interpretation and report, obtained without diagnostic ultrasound examination of the same anatomy (eg, organ, gland, tissue, target structure)
0690T Quantitative ultrasound tissue characterization (non-elastographic), including interpretation and report, obtained with diagnostic ultrasound examination of the same anatomy (eg, organ, gland, tissue, target structure) (List separately in addition to code for primary procedure)
76391 Magnetic resonance (eg, vibration) elastography
76981 Ultrasound, elastography; parenchyma (eg, organ)
76982 - 76983 Ultrasound, elastography

Other CPT codes related to the CPB:

19301 Mastectomy, partial (eg, lumpectomy, tylectomy, quadrantectomy, segmentectomy);
19302     with axillary lyphadenectomy
77065-77067 Screening mammography

HCPCS codes not covered for indications listed in the CPB:

C9788 Opto-acoustic imaging, breast (including axilla when performed), unilateral, with image documentation, analysis and report, obtained with ultrasound examination

ICD-10 codes not covered for indications listed in the CPB:

C50.011 - C50.929 Malignant neoplasm of breast
C79.81 Secondary malignant neoplasm of the breast
D05.00 - D05.92 Carcinoma in situ of breast
D24.1 - D24.9 Benign neoplasm of breast
N60.01 - N65.1 Disorders of breast
Z12.39 Encounter for other screening for malignant neoplasm of breast

Background

Mammography remains the generally accepted standard for breast cancer screening and diagnosis.  However, efforts to provide new insights regarding the origins of breast disease and to find different approaches for addressing several key challenges in breast cancer, including detecting disease in mammographically dense tissue, distinguishing between malignant and benign lesions, and understanding the impact of neoadjuvant chemotherapies, has lead to the investigation of several novel methods of breast imaging for breast cancer management.

Breast Transillumination

Optical imaging involves passing light through the breast, or reflecting light off of the breast, and then measuring the light that returns. It is purportedly useful in the detection of tumors or the blood vessels that supply them. Breast transillumination (also known as light scanning, diaphanography, optical imaging, optical mammography, dynamic optical breast imaging (DOBI), optical transillumination spectroscopy, diffuse optical spectroscopy (DOS), and transillumination breast spectroscopy) is based on the theory that normal and abnormal tissues reflect different light intensities.  Various transilluminators have been created to screen and diagnose cancers of the breast.  One such device, the Computed Tomography Laser Mammography (CTLM) (Imaging Diagnostic Systems Inc., Fort Lauderdale, FL), is a near-infrared (NIR) laser breast imaging system that measures the absorption of NIR light by hemoglobin.  Another method currently under investigation is opto-acoustic tomography (OAT) imaging.  This hybrid imaging technique combines optic and acoustic imaging to map the distribution of optical absorption within biological tissues by means of laser-induced ultrasonic signals.

Breast transilluminators have not been proven to be an acceptable alternative to conventional mammography.  The U.S. Food and Drug Administration (FDA)'s Obstetrics and Gynecology Devices Advisory Panel (1991) considered the clinical utility of breast transilluminators and concluded that "[e]xcept in investigational settings, the devices do not provide meaningful clinical information and should not be used in the clinical evaluation of breast tissue, neither alone nor in conjunction with other techniques."  An Agency for Healthcare Research and Quality Clinical Practice Guideline (Bassett et al, 1994) concluded that “[l]ight scanning (diaphanography and transillumination) should not be used for screening or diagnostic evaluation of the breast”.  The British Columbia Cancer Agency (2001) has concluded that “[a]t the present state of development, transillumination is not an appropriate imaging device for breast cancer screening.”  The American Cancer Society (2009) stated that these experimental imaging tests are still in the earliest stages of research.

Electrical Impedance Scanning (EIS)

Experimental studies have shown that significant changes occur in the electrical properties of breast cancer tissue compared to the surrounding normal tissue.  This phenomenon motivated studies on cancer detection using electrical impedance techniques.  Some evidence has been found that malignant breast tumors have lower electrical impedance than surrounding normal tissues.  This observation has led to the proposal that electrical impedance could be used as an indicator for breast cancer detection.  However, the separation of malignant tumors from benign lesions based on impedance measurements needs further investigation. 

Electrical impedance scanning (EIS) involves continuous transmission of electricity into the body via an electrical patch on the arm or a handheld device. As the current travels through the breast, it is measured by a probe on the surface of the skin. Cancerous and normal tissue conduct electricity differently, and cancerous images purportedly appear as bright white areas.

There are no prospective clinical studies demonstrating the clinical utility of electrical impedance scanning (EIS) in distinguishing benign from malignant breast lesions, either in place of or as an adjunct to mammography or magnetic resonance imaging.  An assessment of technologies for breast cancer screening and diagnosis conducted by the Institute of Medicine of the National Academy of Sciences (2001) concluded that “[c]linical data suggest the technology [EIS] could play a role in breast cancer detection, but more study is needed to define a role in relation to existing technologies.”

Stojadinovic et al (2005) presented preliminary results on the use of EIS for the early detection of breast cancer in young women.  They stated that EIS appears promising for early detection of breast cancer, and identification of young women at increased risk for having the disease at time of screening.  Positive EIS-associated breast cancer risk compares favorably with relative risks of conditions commonly used to justify early breast cancer screening.  The authors also noted that more data are needed to ascertain more accurately the actual sensitivity.  These investigators also believe that EIS has promise as a breast cancer screening modality for a group of women for whom no effective screening modality currently exists.  Electrical impedance scanning seems to identify a population at increased risk for having breast cancer for whom further imaging examinations may be warranted.

On August 29, 2006, the FDA's Obstetrics and Gynecology Devices Panel voted unanimously not to recommend approval of Mirabel Medical Systems' T-Scan 2000 ED bioimpedance device, which is designed to evaluate the risk of breast cancer in asymptomatic women aged 30 to 39 years with no family history of breast cancer and no other known risk factors.  The device would be employed in combination with clinical breast examination for this age group whose annual examination does not usually entail mammography.

The FDA panel decided that the data did not provide a reasonable assurance of the effectiveness to support the device's proposed indication.  Furthermore, some panel members were concerned with other aspects of the clinical trial:
  1. the apparent differences in the characteristics of the 2 trial populations (1,751 women aged 30 to 39 years in the study arm designed to measure specificity and 390 women aged 30 to 45 years in the study arm measuring sensitivity),
  2. a lack of ethnicity data, and
  3. a "high" false-positive rate (Taulbee, 2006).

Blackmore et al (2007) stated that risk assessment by parenchymal density pattern, a strong physical indicator of future breast cancer risk, is available with the onset of mammographic screening programmes.  However, due to the use of ionizing radiation, mammography is not recommended for use in younger women, thereby rendering risk assessment unattainable at an earlier age.  These investigators reported on the use of visible and near infra-red light on 292 women with radiologically normal mammograms to determine if transillumination breast spectroscopy (TIBS) can identify women with a high parenchymal density pattern as an intermediate indicator of breast cancer risk.  Principal component analysis was used to reduce the spectral data and generate density scores for each woman.  To assess the accuracy of TIBS, logistic regression was used to calculate crude and adjusted odds ratios (OR) and 95 % confidence intervals (CI) for each score.  Receiver operator characteristic curves and area under the curve (AUC) were also calculated for the crude and adjusted logistic models.  Optical information relating to tissue chromophores, such as water, lipid and hemoglobin content, was sufficient to identify women with high parenchymal density.  The resulting AUC for the final and most parsimonious multi-variate logistic model was 0.922 (95 % CI: 0.878 to 0.967).  The authors concluded that TIBS provides information correlating to high parenchymal density and is a promising tool for risk assessment, particularly for younger women.  Furthermore, Blackmore and colleagues (2008) also reported that TIBS scores may prove useful as intermediate markers in studies of breast cancer etiology and prevention.

In a prospective, two-cohort trial of pre-menopausal women, Stojadinovic et al (2008) estimated the relative probability of breast cancer in T-Scan+ women compared to randomly selected young women.  The Specificity (S(p))-Cohort evaluated T-Scan specificity in 1,751 asymptomatic women aged 30 to 39.  The Sensitivity)S(n))-Cohort evaluated T-Scan sensitivity in 390 women aged 45 to 30 scheduled for biopsy.  Specificity, sensitivity, and conservative estimate of disease prevalence were used to calculate relative probability.  In the S(p)-Cohort, 93 of 1,751 women were T-Scan+ (S(p) = 94.7 %; 95 % CI: 93.7 to 95.7 %).  In the S(n)-Cohort, 23 of 87 biopsy-proven cancers were T-Scan+ (S(n) = 26.4 %; 95 % CI: 17.4 to 35.4 %).  Given S(p) = 94.7 %, S(n) = 26.4 % and prevalence of 1.5 cancers/1,000 women (aged 30 to 39), the relative probability of a T-Scan+ woman having Br-Ca is 4.95: (95 % CI: 3.16 to 7.14).  The authors concluded that EIS can identify a subset of young women with a relative probability of breast cancer almost 5 times greater than in the population of young women at-large.  The drawbacks of this study were discussed by the afore-mentioned FDA panel.

In a prospective, multi-center study, Wang et al (2010) reported the sensitivity and specificity for the combination of EIS and ultrasound in identifying breast cancer and calculated the relative risk of breast cancer in young women (n = 583) aged 45 years and under scheduled for mammary biopsy.  Of the 583 cases, 143 were diagnosed with breast cancer.  The sensitivities of EIS, ultrasound and the combination method were 86.7 % (124/143), 72 % (103/143), and 93.7 % (134/143) and the specificities were 72.9 % (321/440), 82.5 % (363/440), and 64.1 % (282/440), respectively.  The relative possibilities of breast cancer for the positive young women detected by EIS, ultrasound, and the combination method were 8.67, 5.77, and 14.84, respectively.  The authors concluded that the combination of EIS and ultrasound is likely to become an applicable method for early detection of breast cancer in young women. 

Elastography / Ultrasound Elastography

Elastography refers to the measurement of elastic properties of tissues and is based on the principle that malignant tissue is harder than benign tissue.  Manual palpation in the detection of breast cancer suggests that breast elastography could potentially provide a diagnostic tool for detecting cancerous lesions deeper within the breast.  The technique is typically performed with ultrasound (US), but research with magnetic resonance (MR) is also under way.  Advantages of the US elastography are ubiquitous applicability and cost-effectiveness.  Magnetic resonance elastography (MRE) offers improved reconstruction and the possibility to assess potential anisotropic properties. 

There are 3 main types of US elasticity imaging:
  1. elastography that tracks tissue movement during compression to obtain an estimate of strain,
  2. sonoelastography that uses color Doppler to generate an image of tissue movement in response to external vibrations, and
  3. tracking of shear wave propagation through tissue to obtain the elastic modulus.

The SonixTouch Ultrasound Imaging System received 510(k) marketing clearance by the FDA in October, 2008.  The device includes an elastography imaging mode.  Mechanical pressure with the hand on the transducer produces an imaging sequence similar to the B-mode sequence except the system will acquire the radio-frequency signal instead of acquiring B mode data.  The algorithm extracts a strain value information for every point on the image.  The elastography image then color-codes the stiff versus softer structures.

Magnetic resonance elastography is a phase-contrast-based magnetic resonance imaging (MRI) technique that can directly visualize and quantitatively measure propagating acoustic strain waves in tissue subjected to harmonic mechanical excitation.  The data acquired allows the calculation of local quantitative values of shear modulus and the generation of images that depict tissue elasticity or stiffness.

Garra et al (1997) examined the feasibility of elastography to determine the appearance of various breast lesions and the potential of elastography to diagnosis breast lesions.  A total of 46 breast lesions were examined with elastography.  Patients underwent biopsy or aspiration of all lesions, revealing 15 fibroadenomas, 12 carcinomas, 6 fibrocystic nodules, and 13 other lesions.  The elastogram was generated from radio-frequency data collected with use of a 5-MHz linear-array transducer.  The elastogram and corresponding sonogram were evaluated by a single observer for lesion visualization, relative brightness, and margin definition and regularity.  The sizes of the lesions at each imaging examination and at biopsy were recorded and compared.  Softer tissues such as fat appeared as bright areas on elastograms.  Firm tissues, including parenchyma, cancers, and other masses, appeared darker.  The cancers were statistically significantly darker than fibroadenomas (p < 0.005) and substantially larger on the elastogram than on the sonogram.  A total of 73 % of fibroadenomas and 56 % of solid benign lesions could be distinguished from cancers by using lesion brightness and size difference.  Some cancers that appeared as areas of shadowing on sonograms appeared as discrete masses on elastograms.  The authors concluded that elastography has the potential to be useful in the evaluation of areas of shadowing on the sonogram and that it may be helpful in the distinction of benign from malignant masses.

Lorenzen et al (2002) explored the potential of elasticity as a parameter for the diagnosis of breast lesions using MRE in 15 female patients with malignant tumors of the breast, 5 patients with benign breast tumors, and 15 healthy volunteers.  Malignant invasive breast tumors documented the highest values of elasticity with a median of 15.9 kPa and a wide range of stiffnesses between 8 and 28 kPa.  In contrast, benign breast lesions represented low values of elasticity, which were significantly different from malignant breast tumors (median elasticity: 7.0 kPa; p = 0.0012).  This was comparable to the stiffest tissue areas in healthy volunteers (median elasticity 7.0 kPa), whereas breast parenchyma (median: 2.5 kPa) and fatty breast tissue (median: 1.7 kPa) showed the lowest values of elasticity. Two invasive ductal carcinomas had elasticity values of 8 kPa and two stiff parenchyma areas in healthy volunteers had elasticities of 13 and 15 kPa.  These lesions could not be differentiated by their elasticity.  The authors concluded that MRE is a promising new imaging modality with the capability to assess the viscoelastic properties of breast tumors and the surrounding tissues; however, there is an overlap in the elasticity ranges of soft malignant tumors and stiff benign lesions.

Sinkus et al (2005) examined the viscoelastic shear properties of breast lesions measured by MRE applied in the course of standard MR mammography to 15 patients with different pathologies (6 breast cancer cases, 6 fibroadenoma cases and 3 mastopathy cases).  Breast cancer appeared on average 2.2 (p < 0.001) times stiffer.  All breast cancer cases showed a good delineation to the surrounding breast tissue with an average elevation of a factor of 3.3 (p < 1.4 x 10(-6)).  However, the results were not found to be useful for separating benign from malignant lesions.

Giuseppetti et al (2005) assessed the diagnostic accuracy of elastography in characterizing nodular breast lesions.  A total of 82 patients who received mammographic, US and elastographic evaluation in a single session at two Italian centers between January and August 2004 according to identical protocols exhibited 91 nodules that were subjected to cytological/histological examination.  Lesions were classified and scored and the sensitivity and specificity of elastography calculated.  Overall sensitivity and specificity were 79 % and 89 %, respectively.  However, sensitivity was 86 % and 65 % and specificity 100 % and 62 % for lesions less than 2 cm and greater than 2 cm in diameter, respectively.  The authors concluded that larger studies are needed to establish semiological patterns.

Itoh et al (2006) evaluated the diagnostic performance of real-time elastography (RTE) by using the extended combined autocorrelation method (CAM) to differentiate benign from malignant breast lesions, with pathologic diagnosis as the reference standard.  Conventional US and RTE with CAM were performed in 111 women (mean age of 49.4 years; age range of 27 to 91 years) who had breast lesions (59 benign, 52 malignant).  Elasticity images were assigned an elasticity score according to the degree and distribution of strain induced by light compression.  The area under the curve and cut-off point, both of which were obtained by using a receiver operating characteristic curve analysis, were used to assess diagnostic performance.  Mean scores were examined by using a Student t test.  Sensitivity, specificity, and accuracy were compared by using the standard proportion difference test or the Delta-equivalent test.  For elasticity score, the mean +/- standard deviation was 4.2 +/-  0.9 for malignant lesions and 2.1 +/- 1.0 for benign lesions (p < 0.001). When a cut-off point of between 3 and 4 was used, elastography had 86.5 % sensitivity, 89.8 % specificity, and 88.3 % accuracy.  When a best cut-off point of between 4 and 5 was used, conventional US had 71.2 % sensitivity, 96.6 % specificity, and 84.7 % accuracy.  Elastography had higher sensitivity than conventional US (p < 0 .05).  By using equivalence bands for non-inferiority or equivalence, it was shown that the specificity of elastography was not inferior to that of conventional US and that the accuracy of elastography was equivalent to that of conventional US.  The authors concluded that US elastography with the proposed imaging classification had almost the same diagnostic performance as conventional US.

Thomas et al (2006) compared the sensitivity and specificity of elastography with that of B-mode US and mammography in 300 patients with histologically confirmed breast lesions (168 benign, 132 malignant).  Evaluation was by means of the 3-dimensional (3-D) finite-element method.  The data were color-coded and superimposed on the B-mode US scan.  The images were evaluated by two independent readers.  The results were compared with mammography, histology, and the data obtained by previous US investigations.  Sensitivity and specificity in the differentiation of benign and malignant lesions were 87 % and 85 %, respectively, for mammography and 94 % and 83 % for B-mode US.  The 2 examiners were in very good agreement in their evaluation of the elastograms (kappa: 0.86).  Elastography had a sensitivity of 82 % and a specificity of 87 %.  Elastography was superior to B-mode US in diagnosing Breast Imaging Reporting and Data System (BI-RADS) 3 lesions (92 % versus 82 % specificity) and in lipomatous involution (80 % versus 69 % specificity).

Zhang et al (2006) investigated the clinical value of RTE in the diagnosis of breast cancer in 120 patients with breast lumps (135 lesions).  Patients were examined with B-mode imaging, color Doppler flowing imaging (CDFI) and RTE.  The elastogram was graded using a 5-score evaluating method.  The post-operative pathological diagnosis was used as the gold standard, and the sensitivity, specificity and accuracy of RTE and 2-D US combined with RTE in diagnosis of breast cancer were calculated.  When the score greater than 4 was set for cut-off criteria of malignancy, the sensitivity, specificity and accuracy of RTE was 85.45 %, 83.75 % and 84.4 %, respectively.  When 2-D US combined with RTE was used, the sensitivity, specificity and accuracy increased up to 100 %, 95 % and 97 %, respectively.

Thomas et al (2006) evaluated whether RTE could improve the differentiation and characterization of benign and malignant breast lesions.  Real-time elastography was carried out in 108 potential breast tumor patients with cytologically/histologically confirmed focal breast lesions (59 benign, 49 malignant; median age, 53.9 years; range of 16 to 84 years).  Tumor and healthy tissue were differentiated by measurement of elasticity based on the correlation between tissue properties and elasticity modulus.  Evaluation was performed using the 3-D finite element method, in which the information is color-coded and superimposed on the B-mode US image.  A second observer evaluated the elastography images, in order to improve the objectivity of the method.  The results of B-mode scan and elastography were compared with those of histology and previous sonographic findings.  Sensitivities and specificities were calculated, using histology as the gold standard.  B-mode US had a sensitivity of 91.8 % and a specificity of 78 %, compared with sensitivities of 77.6 % and 79.6 % and specificities of 91.5 % and 84.7 %, respectively, for the two observers evaluating elastography.  Agreement between B-mode US and elastography was good, yielding a weighted kappa of 0.67.  The authors concluded that RTE improves the specificity of breast lesion diagnosis and is a promising new approach for the diagnosis of breast cancer. 

Zhi et al (2007) compared the use of US elastography with mammography, and sonography in the diagnosis of solid breast lesions.  From September 2004 to May 2005, 296 solid lesions from 232 consecutive patients were diagnosed as benign or malignant by mammography and sonography and further analyzed with US elastography.  The diagnostic results were compared with histopathologic findings.  The sensitivity, specificity, accuracy, positive and negative predictive values, and false-positive and -negative rates were calculated for each modality and the combination of US elastography and sonography.  Of 296 lesions, 87 were histologically malignant, and 209 were benign.  Ultrasound elastography was the most specific (95.7 %) and had the lowest false-positive rate (4.3 %) of the 3 modalities.  The accuracy (88.2 %) and positive predictive value (PPV) (87.1 %) of US elastography were higher than those of sonography (72.6 % and 52.5 %, respectively).  The sensitivity values, negative predictive values (NPV), and false negative rates of the 3 modalities had no differences.  A combination of US elastography and sonography had the best sensitivity (89.7 %) and accuracy (93.9 %) and the lowest false-negative rate (9.2 %).  The specificity (95.7%) and PPV (89.7%) of the combination were better, and the false-positive rate (4.3%) of the combination was lower than those of mammography and sonography.  The authors concluded that US elastography is a promising technique for evaluating breast lesions.

Tardivon et al (2007) evaluated elastography in the characterization of breast nodules in 122 lesions.  Elastography (Hitachi, 7.5- to 13-MHz probe; Ueno classification, scores 1-3 = benign, 4-5 = malignant) was evaluated in 125 sub-clinical lesions in 114 patients.  The results were compared to those of the American College of Radiology's BI-RADS sonography categories (benign = 2 and 3, malignant = 4 and 5) and to the results of the percutaneous samples taken and/or surgery (122 lesions evaluated, 59 % less than 10 mm, 61 cancers, 61 benign lesions).  There were 3 technical failures (2.4 %).  The elastography was in agreement with histology for 101 lesions, with 13 false-negative results and 8 false-positive results (sensitivity, 78.7 %; specificity, 86.9 %; PPV, 85.7 %; NPV, 80.3 %); versus agreement with the BI-RADS classification for 98 lesions with one false-negative result and 23 false-positive results (sensitivity, 98.4 %; specificity, 47.5 %; PPV, 65.2 %; NPV, 96.7 %).

Cho et al (2008) compared the diagnostic performances of conventional US and US elastography for the differentiation of non-palpable breast masses, and to evaluate whether elastography is helpful at reducing the number of benign biopsies, using histological analysis as a reference standard.  Conventional US and RTE images were obtained for 100 women who had been scheduled for a US-guided core biopsy of 100 non-palpable breast masses (83 benign, 17 malignant).  Two experienced radiologists unaware of the biopsy and clinical findings analyzed conventional US and elastographic images by consensus, and classified lesions based on degree of suspicion regarding the probability of malignancy.  Results were evaluated by receiver operating characteristic (ROC) curve analysis.  In addition, the authors investigated whether a subset of lesions were categorized as suspicious by conventional US, but as benign by elastography.  Areas under the ROC curves (Az values) were 0.901 for conventional US and 0.916 for elastography (p = 0.808).  For BI-RADS category 4a lesions, 44 % (22 of 50) had an elasticity score of 1 and all were found to be benign.  The authors concluded that elastography had a diagnostic performance comparable to that of conventional US for the differentiation of non-palpable breast masses.

Researchers have tested the feasibility of breast elastography and the results confirm the hypothesis that breast elastography can quantitatively depict the elastic properties of breast tissues and reveal high shear elasticity in known breast tumors.  However, the clinical benefits of elastography imaging are still under evaluation and no clinical diagnosis can be made other than being able to tell whether or not a structure inside the patient is stiffer than another one.  Further research is needed to evaluate the potential clinical applications of breast elastography, such as detecting breast carcinoma and characterizing suspicious breast lesions.

Kumm and Szabunio (2010) evaluated the application and diagnostic performance of elastography for the characterization of breast lesions in patients referred for biopsy.  Subjects referred for ultrasound-guided biopsy of sonographically apparent breast lesions were included in this study.  The Hitachi Hi-Vision 900 ultrasound was used to generate index test results for elastography scoring (ES) and for strain ratio (SR) measurement.  Sensitivity, specificity, PPV and NPV were determined using pathologic results from 14-gauge core needle biopsy as the reference standard.  A total of 310 lesions in 288 patients were included in this series.  Of these 310 lesions, 223 (72 %) were benign and 87 (28 %) were malignant.  Sensitivity was 0.76 for ES and 0.79 for SR.  Specificity was 0.81 for ES and 0.76 for SR.  The PPV was 0.60 for ES and 0.57 for SR; the NPV was 0.90 for ES and 0.90 for SR.  Strain ratio values for malignant lesions were significantly higher (median ratios of 10.5 and 2.7, respectively, p < 0.001).  The authors concluded that while the initial clinical performance of elastography imaging shows potential to reduce biopsy of low-risk lesions, a large-scale trial addressing appropriate patient selection, diagnostic parameters, and practical application of this technique is needed before widespread clinical use.

Siegmann et al (2010) assessed the additional value of MRE to contrast-enhanced (ce) MRI for breast lesion characterisation.  A total fo 57 suspected breast lesions in 57 patients (mean age of 52.4 years) were examined by ce MRI and MRE.  All lesions were classified into BI-RADS categories.  Viscoelastic parameters, e.g., alpha0 as an indicator of tissue stiffness, were calculated. Histology of the lesions was correlated with BI-RADS and viscoelastic properties.  The PPV for malignancy, and the sensitivity and specificity of ce MRI were calculated.  The ROC curves were separately calculated for both ce MRI and viscoelastic properties and conjoined to analyse the accuracy of diagnostic performance.  The lesions (mean size of 27.6 mm) were malignant in 64.9 % (n = 37) of cases.  The PPV for malignancy was significantly (p < 0.0001) dependent on BI-RADS classification.  The sensitivity of ce MRI for breast cancer detection was 97.3 % (36/37), whereas specificity was 55 % (11/20).  If ce MRI was combined with alpha0, the diagnostic accuracy could be significantly increased (p < 0.05; AUC(ce MRI) = 0.93, AUC(combined) = 0.96).  The authors concluded that the combination of MRE and ce MRI could increase the diagnostic performance of breast MRI.  Moreover, they stated that further investigations of larger cohorts and smaller lesions (in particular those only visible on MRI) are needed to validate these findings.

Yi et al (2012) evaluated the diagnostic value of sonoelastography by correlation with histopathology compared with conventional ultrasound on the decision to biopsy.  Prospectively determined BI-RADS categories of conventional ultrasound and elasticity scores from strain sonoelastography of 1,786 non-palpable breast masses (1,523 benign and 263 malignant) in 1,538 women were correlated with histopathology.  The sensitivity and specificity of 2 imaging techniques were compared regarding the decision to biopsy.  These researchers also investigated whether there was a subset of benign masses that were recommended for biopsy by B-mode ultrasound but that had a less than 2 % malignancy rate with the addition of sonoelastography.  The mean elasticity score of malignant lesions was higher than that of benign lesions (2.94 +/- 1.10 versus 1.78 +/- 0.81) (p < 0.001).  In the decision to biopsy, B-mode ultrasound had higher sensitivity than sonoelastography (98.5 % versus 93.2 %) (p < 0.001), whereas sonoelastography had higher specificity than B-mode ultrasound (42.6 % versus 16.3 %) (p < 0.001).  BI-RADS category 4a lesions with an elasticity score of 1 had a malignancy rate of 0.8 %.  The authors concluded that sonoelastography has higher specificity than B-mode ultrasound in the differentiation between benign and malignant masses and has the potential to reduce biopsies with benign results.

Landoni and colleagues (2012) developed a quantitative method for breast cancer diagnosis based on elastosonography images in order to reduce whenever possible unnecessary biopsies.  The proposed method was validated by correlating the results of quantitative analysis with the diagnosis assessed by histopathologic examination.  A total of 109 images of breast lesions (50 benign and 59 malignant) were acquired with the traditional B-mode technique and with elastographic modality.  Images in Digital Imaging and COmmunications in Medicine format (DICOM) were exported into a software, written in Visual Basic, especially developed to perform this study.  The lesion was contoured and the mean grey value and softness inside the region of interest (ROI) were calculated.  The correlations between variables were investigated and receiver operating characteristic (ROC) curve analysis was performed to assess the diagnostic accuracy of the proposed method.  Pathologic results were used as standard reference.  Both the mean grey value and the softness inside the ROI resulted statistically different at the t-test for the 2 populations of lesions (i.e., benign versus malignant): p <0.0001.  The area under the curve (AUC) was 0.924 (0.834 to 0.973) and 0.917 (0.826 to 0.970) for the mean grey value and for the softness respectively.  The authors concluded that quantitative elastosonography is a promising ultrasound technique in the detection of breast cancer; but large prospective trials are needed to determine if quantitative analysis of images can help to overcome some pitfalls of this method.

In a meta-analysis, Sadigh et al (2012) reviewed evidence on diagnostic performance of strain ratio and length ratio, 2 different strain measurements in ultrasound elastography (USE), for differentiating benign and malignant breast masses.  A literature search of PubMed and other medical and general purpose databases from inception through January 2012 was conducted.  Published studies that evaluated the diagnostic performance of USE alone reporting either strain ratio or length ratio for characterization of focal breast lesions and using cytology (fine needle aspiration) or histology (core biopsy) as a reference standard were included.  Summary diagnostic performance measures were assessed using bi-variate generalized linear mixed modeling.  A total of 9 studies reported strain ratio for 2,087 breast masses (667 cancers, 1,420 benign lesions).  Summary sensitivity and specificity were 88 % (95 % CI: 84 to 91 %), and 83 % (95 % CI: 78-88 %), respectively.  The positive and negative likelihood ratios (LR) were 5.57 (95 % CI: 3.85 to 8.01) and 0.14 (95 % I: 0.09 to 0.20), respectively.  The inconsistency index for heterogeneity was 6 % (95 % CI: 1 to 22 %) for sensitivity and 8 % (95 % CI: 3 to 24 %) for specificity.  Analysis of 3 studies reporting length ratio for 450 breast masses demonstrated sensitivity and specificity of 98 % (95 % CI: 93 to 99 %) and 72 % (95 % CI: 31 to 96 %), respectively.  Strain ratio and length ratio have good diagnostic performance for distinguishing benign from malignant breast masses.  The authors concluded that although, this performance may not be incrementally superior to that of BI-RADS in B-mode ultrasound, the application of USE using strain ratio or length ratio in combination with USB may have the potential to benefit the patients, and this requires further comparative effectiveness and cost-effectiveness analyses.

Vreugdenburg et al (2013) evaluated all the available evidence of safety, effectiveness and diagnostic accuracy for three emerging classes of technology promoted for breast cancer screening and diagnosis: Digital infrared thermal imaging (DITI), EIS and elastography.  A systematic search of seven biomedical databases (EMBASE, PubMed, Web of Science, CRD, CINAHL, Cochrane Library, Current Contents Connect) was conducted through March 2011, along with a manual search of reference lists from relevant studies.  The principal outcome measures were safety, effectiveness, and diagnostic accuracy.  Data were extracted using a standardized form, and validated for accuracy by the secondary authors.  Study quality was appraised using the quality assessment of diagnostic accuracy studies tool, while heterogeneity was assessed using forest plots, Cooks' distance and standardized residual scatter plots, and I (2) statistics.  From 6,808 search results, a total of 267 full-text articles were assessed, of which 60 satisfied the inclusion criteria.  No effectiveness studies were identified.  Only 1 EIS screening accuracy study was identified, while all other studies involved symptomatic populations.  Significant heterogeneity was present among all device classes, limiting the potential for meta-analyses.  Sensitivity and specificity varied greatly for DITI (Sens = 0.25 to 0.97, Spec = 0.12 to 0.85), EIS (Sens = 0.26 to 0.98, Spec = 0.08-to 0.81) and ultrasound elastography (Sens = 0.35 to 1.00, Spec = 0.21 to 0.99).  The authors concluded that there is currently insufficient evidence to recommend the use of these technologies for breast cancer screening.  Moreover, the high level of heterogeneity among studies of symptomatic women limits inferences that may be drawn regarding their use as diagnostic tools.  They stated that future research employing standardized imaging, research and reporting methods is needed.

Olgun and colleagues (2014) determined the correlations between the elasticity values of solid breast masses and histopathological findings to define cut-off elasticity values differentiating malignant from benign lesions.  A total of 115 solid breast lesions of 109 consecutive patients were evaluated prospectively using shear wave elastography (SWE).  Two orthogonal elastographic images of each lesion were obtained.  Minimum, mean, and maximum elasticity values were calculated in regions of interest placed over the stiffest areas on the 2 images; these researchers also calculated mass/fat elasticity ratios.  Correlation of elastographic measurements with histopathological results were studied.  A total of 83 benign and 32 malignant lesions were histopathologically diagnosed.  The minimum, mean, and maximum elasticity values, and the mass/fat elasticity ratios of malignant lesions, were significantly higher than those of benign lesions.  The cut-off value was 45.7 kPa for mean elasticity (sensitivity, 96 %; specificity, 95 %), 54.3 kPa for maximum elasticity (sensitivity, 95 %; specificity, 94 %), 37.1 kPa for minimum elasticity (sensitivity, 96 %; specificity, 95 %), and 4.6 for the mass/fat elasticity ratio (sensitivity, 97 %; specificity, 95 %).  The authors concluded that WE yielded additional valuable quantitative data to ultrasonographic examination on solid breast lesions; and SWE may serve as a complementary tool for diagnosis of breast lesions.  Moreover, they stated that long-term clinical studies are needed to accurately select lesions requiring biopsy.

In a meta-analysis, Chen and colleagues (2014) examined the performance of SWE for the differentiation of benign and malignant breast lesions. PubMed, Embase and the Cochrane library were searched for studies published up to January 2014. The references of retrieved relevant articles were reviewed to identify potential publications. Random-effect meta-analysis was conducted to assess the overall sensitivity and specificity of SWE in the differentiation of breast lesions. A total of 11 articles, including 2,424 patients, were included in the present meta-analysis. The summarized sensitivity and specificity of the SWE performance based on maximum elasticity were 0.93 (95 % CI: 0.9 to 0.95) and 0.81 (95 % CI: 0.78 to 0.83), respectively. For the mean elasticity, the summarized sensitivity and specificity were 0.94 (95 % CI: 0.92 to 0.96) and 0.71 (95 % CI: 0.69 to 0.74), respectively. The summarized sensitivity and specificity were 0.77 (95 % CI: 0.70 to 0.83) and 0.88 (95 % CI: 0.84 to 0.91) for the SD of elasticity. The authors concluded that SWE has a high sensitivity and specificity in the differentiation of benign and malignant breast lesions. However, they stated that more large and prospective studies are needed to further examine the performance of SWE.

Chamming's et al (2015) evaluated imaging performances for the detection, characterization and biopsy of breast micro-calcifications and made recommendations.  French and English publications were searched using PubMed, Cochrane Library and international learned societies recommendations.  Digital mammography (DR [Direct Radiography] and CR [Computed Radiography]) and screen-film mammography demonstrated good performances for the detection and the characterization of breast micro-calcifications.  Systematic use of the 2013 edition of the BI-RADS lexicon was recommended for description and characterization of micro-calcifications.  Faced with BI-RADS 4 or 5 micro-calcifications, breast ultrasound is recommended but a normal result does not eliminate the diagnosis of cancer and other examination should be performed.  Literature review does not allow recommending digital breast tomosynthesis, elastography or MRI to analyze micro-calcifications.  In case of probably benign micro-calcifications (BI-RADS 3), 6 months, 1 year and at least 2 years follow-up are recommended.  In case a biopsy is indicated, it is recommended to use a vacuum-assisted macro-biopsy system with 11-G needles or bigger.  If no calcification is visible on the radiography of the specimen, it is recommended to obtain additional samples.

In a meta-analysis, Liu et al (2016) estimated the diagnostic performance of SWE in differentiating malignant from benign breast lesions.  A literature search of PubMed, Web of Science and Scopus up to November 2014 was conducted.  A summary receiver operating characteristic curve was constructed, and pooled weighted estimates of sensitivity and specificity were calculated using a bi-variate mixed-effects regression model.  A total of 33 studies, which included a total of 5,838 lesions (2,093 malignant, 3,745 benign) from 5,397 patients, were finally analyzed.  Summary sensitivity and specificity were 0.886 (95 % CI: 0.858 to 0.909) and 0.866 (95 % CI: 0.833 to 0.894), respectively.  The pooled diagnostic OR was 50.41 (95 % CI: 34.972 to 72.664).  And the area under the receiver operating characteristic curve of SWE was 0.94 (95 % CI: 0.91 to 0.96).  No publication bias existed among these studies (p = 0.245).  In the subgroup analysis, sensitivity and specificity were 0.862 (95 % CI: 0.811 to 0.901) and 0.875 (95 % CI: 0.793 to 0.928) among 1,552 lesions from 1,429 patients in the 12 studies using acoustic radiation force impulse imaging and 0.897 (95 % CI: 0.863 to 0.923) and 0.863 (95 % CI: 0.831 to 0.889) among another 4,436 lesions from 4,097 patients in the 21 studies using supersonic shear imaging.  When analysis confined to 9 studies evaluated the diagnostic performance of combination SWE and conventional ultrasound, the area under the curve was 0.96 (95 % CI: 0.94 to 0.97), yielding a sensitivity of 0.971 (95 % CI: 0.941 to 0.986) and specificity of 0.801 (95 % CI: 0.733 to 0.856).   The authors concluded that SWE appeared to be a good quantitative method for differentiating breast lesions, with promise for integration into routine imaging protocols.

In a retrospective study, Chung and associates (2016) evaluated the diagnostic performance of SWE for the differential diagnosis of breast papillary lesions.  A total of 79 breast papillary lesions in 71 consecutive women underwent ultrasound and SWE prior to biopsy.  Ultrasound features and quantitative SWE parameters were recorded for each lesion.  All lesions were surgically excised or excised using an ultrasound-guided vacuum-assisted method.  The diagnostic performances of the quantitative SWE parameters were compared using the AUC.  Of the 79 lesions, 6 (7.6 %) were malignant and 12 (15.2 %) were atypical.  Orientation, margin, and the final BI-RADS ultrasound assessments were significantly different for the papillary lesions (p < 0.05).  All qualitative SWE parameters were significantly different (p < 0.05).  The AUC values for SWE parameters of benign and atypical or malignant papillary lesions ranged from 0.707 to 0.757 (sensitivity, 44.4 to 94.4 %; specificity, 42.6 to 88.5 %).  The maximum elasticity and the mean elasticity showed the highest AUC (0.757) to differentiate papillary lesions.  The authors concluded that SWE provided additional information for the differential diagnosis of breast papillary lesions; quantitative SWE features were helpful to differentiate breast papillary lesions.  Moreover, they stated that further study about breast papillary lesions using larger ROI size may be necessary.

This study had several drawbacks:
  1. small sample size (n = 71).  Of 217 papillary lesions diagnosed by US-guided CNB, only 79 (36.4 %) were finally included due to the lack of a SWE image or no excision.  Thus, large numbers and multi-center studies are needed in the future,
  2. the authors did not compare their results with those of conventional US or assess combined diagnostic performance; further studies combining SWE and conventional US in are needed to differentiate papillary lesions,
  3. these researchers used a 2x2-mm sized ROI. Larger ROI might be more accurate for the assessment of the breast masses by providing both maximum stiffness and heterogeneity of breast lesions.

In a systematic review and meta-analysis, Blank and colleagues (2017) reported measured elasticity of benign and malignant breast pathologies from SWE, quantitatively confirmed the effect of the selected ROI on these measures and tested the hypothesis that a metric of heterogeneity based on the mean and maximum elasticity can improve specificity of diagnosis.  The elasticity of benign, malignant and specific pathologic states were reported from 22 publications encompassing 2,989 patients, identified from a structured search of the literature from May to September 2015.  A total of 12 articles were included in a meta-analysis that grouped results by the method of ROI selection to discriminate between different pathologies.  These researchers observed a significant correlation between the method of selection of ROI for malignant mean (p < 0.001) and maximum (p = 0.027) elasticity, but no correlation with benign measures.  They defined a quantitative heterogeneity parameter, the "stiffness gradient", computed from the mean and maximum measured elasticity.  The stiffness gradient out-performed the current standard maximum elasticity metric in stratifying malignancy risk by a margin of 15 % for the partial ROI, and 42 % for the maximized ROI.  An anecdotal example of improved differentiation using the stiffness gradient on pathology-specific lesions was also provided.  The authors concluded that these results quantitatively indicated that the method of ROI selection in SWE not only has a significant impact on the resulting mean reported elasticity of a lesion, but may provide some insight into lesion heterogeneity.  They stated that these findings suggested that further exploration of quantitative heterogeneity is needed to improve the specificity of diagnosis.

In a prospective study, Ma and co-workers (2016) evaluated SWE and SWE combined with the Ki-67 index as novel predictive modalities for the pathological response of invasive breast cancer to neoadjuvant chemotherapy (NAC).  This study recruited 66 eligible patients from July 2014 to November 2015.  Tumor stiffness, which corresponds with tumor progression and invasiveness, was assessed by quantitative SWE 1 day before biopsy (time-point t0, elasticity E0), 1 day before next NAC cycle (t1-t5, E1-E5), and 1 day before surgery (t6, E6).  The relative changes in SWE parameters after the 1st and 2nd NAC cycles were considered as the variables [ΔE (t1), ΔE (t2)].  The pathological response was classified according to the residual cancer burden (RCB) protocol.  Correlations between RCB scores and variables were evaluated.  The predictive diagnostic performances of SWE parameters, Ki-67 index, and the predictive RCB (predRCB) score determined by a linear regression model were compared.  Some immunohistochemical and molecular factors and SWE parameters were significantly different among the 3 RCB groups.  The ΔEmean (t2) and Ki-67 had significantly better diagnostic performance than other parameters regarding predicting the pathological response (the RCB-I response and RCB-III resistance).  However, the correlation between ΔEmean (t2) and Ki-67 index was significantly weaker as a diagnostic predictor (r = 0.29).  These researchers generated a new predictive modality, predRCB, which is a multi-variable linear regression model that combines ΔEmean (t2) and the Ki-67 index.  The predRCB modality showed better diagnostic performance than SWE parameters and Ki-67 index alone.  The authors concluded that these findings highlighted the potential utility for adding the Ki-67 index to the SWE results, which may improve the predictive power of SWE and facilitate personalizing the treatment regimens of patients with breast cancer.  Moreover, they stated that these results should be validated in the future by performing a multi-center prospective study with a larger cohort.

Wang and co-workers (2016) identified the automated breast volume scanning (ABVS) and shear wave velocity (SWV) characteristics of different pathological types of breast carcinoma.  These researchers performed a retrospective review of both ABVS and SWV imaging of 118 consecutive breast masses.  The imaging features of both techniques were assessed with reference to histopathological results.  Echo heterogeneity with a smooth and lobulated margin was a significant feature more frequently found in mucinous carcinoma groups (100 %, p < 0.05).  Between different stages of ductal carcinoma, echo homogeneity was more likely in high-grade ductal carcinomas (p < 0.05); SWV differences existed between inside tumor areas and either tumor boundary or tissues outside the tumors (p < 0.05), and values differed between different breast carcinoma stages.  The central and tumor margin areas of ductal carcinomas were much harder than in tubular carcinoma and micro-carcinoma, respectively (p < 0.05); SWV ROC curve analyses yielded a cut-off value of 3.015 m/s between ductal carcinoma in-situ (DCIS) and invasive ductal carcinoma in the central part of lesions, with 83.5 % sensitivity and 80 % specificity for T0 versus T1-3 staging.  The authors concluded that since some features were associated with different breast carcinoma types and stages, ABVS and SWV imaging has the potential to give clues about breast carcinoma differentiation in a non-invasive manner.

de Faria Castro Fleury and co-workers (2017) examined if there is correlation between MRI findings and breast elastography to differentiate seroma/hematoma from silicone-induced granuloma of breast implant capsule (SIGBIC).  This was a prospective study of 99 patients with breast implants submitted to breast MRI during the period from February 1 to May 1, 2017.  Patients who presented with MRI findings of seroma/hematoma or SIGBIC were submitted to a complementary US elastography study to evaluate the correlation of the results.  The criteria adopted for the diagnosis of granuloma by MRI were heterogeneous hyper-signal in the T2-weighted sequences, late contrast enhancement, and black drop sign.  Lesions that did not enhance after the use of contrast were considered as seroma/hematoma.  By elastography, the results were considered positive for granuloma when presented as hard lesions, whereas seroma/hematoma presented as soft lesions.  Of the 99 patients evaluated, 15 were included in the study.  Of the 15 patients, 9 had solid intra-capsular MRI masses, whereas 6 presented collections without contrast enhancement.  The complementary elastography study showed correlation with MRI results in all cases of SIGBIC and seroma/hematoma, with elastography being able to differentiate lesions from solid to cystic.  The authors concluded that elastography of intra-capsular masses in breast implants presented results compatible with those found by MRI for differentiation of solid lesions from collections.  Moreover, the authors stated that this study was a pilot study; these findings must be confirmed by increasing the sample size and performing other multi-center studies to replicate the results.

In a meta-analysis, Xue and colleagues (2017) evaluated the diagnostic accuracy of SWE for malignant breast lesions.  Related articles were searched from PubMed, Embase, and Cochrane library.  Overall sensitivity and specificity were analyzed with DerSimonian and Laird random effects model; AUC with corresponding 95 % CI were calculated to evaluate the diagnostic accuracy of SWE.  Sensitivity and publication bias were assessed as well.  A total of 25 articles including 4,128 patients and 4,546 breast lesions were included in the pooled analysis.  In the subgroup analysis, diagnostic sensitivity and specificity of SWE in Asian population were 0.84 (0.79 to 0.88) and 0.87 (0.84 to 0.90), respectively, whereas they were 0.92 (0.86 to 0.96) and 0.89 (0.84 to 0.92) in Caucasian population.  The diagnostic accuracy of SWE was a little higher for Caucasians than for Asians (0.95 versus 0.92).  The diagnostic sensitivity and specificity of virtual touch tissue quantification were 0.85 (0.77 to 0.91) and 0.93 (0.88 to 0.96), respectively.  It showed a little higher value in specificity and summary ROC curve than SWE (0.93 versus 0.87; 0.95 versus 0.93).  In addition, maximum stiffness exhibited higher detection sensitivity than that of mean stiffness (0.91 versus 0.85).  The authors concluded that with wide application, SWE may significantly improve the early diagnostic of breast cancer.  Moreover, the author noted some limitations of this meta-analysis.  The number of articles based on Caucasian population was much less than that of Asian population.  The accuracy of results on Caucasian population might be affected.  In addition, significant heterogeneity exhibited between the included studies.  The heterogeneity might be caused by the patients’ number, basic feature of patients, and experiments methods, and so on.

In a retrospective study, Kim and co-workers (2017) examined SWE and color Doppler US features for fibro-epithelial lesions (FELs), and evaluated their utility to differentiate fibro-adenomas (FAs) and phyllodes tumors (PTs).  This trial included 67 FELs pathologically confirmed (49 FAs, 18 PTs); B-mode US, SWE and color Doppler US were performed for each lesion.  Mean elasticity (Emean), maximum elasticity (Emax), and vascularity were determined by SWE and Doppler US.  Diagnostic performances were calculated to differentiate FAs and PTs.  Equivocal FELs diagnosed by core needle biopsy (CNB) were further analyzed.  Median Emean and Emax were significantly lower for FAs than PTs (Emean, 15.7 versus 66.7 kPa; Emax, 21.0 versus 76.7 kPa, p < 0.01)).  Low vascularity (0 to 1 vessel flow) on color Doppler US were more frequent in FAs than in PTs (p < 0.01); SWE showed significantly higher specificities (Emean greater than 43.9 kPa, 89.8 %; Emax greater than 46.1 kPa, 79.6 %) than B-mode US (42.9 %) (p < 0.01) for differentiating PTs from FAs.  Other diagnostic values of SWE and overall diagnostic values of Doppler US were not significantly different from B-mode US (p > 0.05).  The combination of SWE and Doppler US with 'Emean greater than 43.9 kPa or high vascularity (greater than or equal to 2 vessel flows)' showed a higher AUC (0.786 versus 0.687) and higher diagnostic values than B-mode US (sensitivity, 100 versus 94.4 %; specificity, 57.1 versus 42.9 %; positive predictive value, 46.2 versus 37.8 %; negative predictive value, 100 versus 95.5 %), without statistical significance (p > 0.05).  Of the 30 equivocal FELs, all lesions with 'Emean less than or equal to 43.9 kPa and low vascularity (0 to 1 vessel flow)' (23.3 %, 7/30) were finally confirmed as FAs by excision.  The authors concluded that FAs had a tendency to have less stiffness and lower vascularity than PTs; combined SWE and color Doppler US may help patients with equivocal FELs diagnosed by CNB avoid unnecessary excision.

The authors stated that this study had several limitations.  First, these researchers could not include all patients who had been diagnosed with FA or PT during the study period; thus, a selection bias might exist.  Since this was a retrospective study using previously static-captured images, lesions which did not have available SWE and color Doppler US data were excluded.  Second, this study had a relatively small sample population, and a further study with a large sample size is needed.  Third, not all tumors were confirmed through total excision and a 1-year follow-up period for FAs without excision was too short to establish benignity.  However, all 24 FAs without total excision were image-pathology concordant and showed stability during each follow-up period (range of 20 to 26 months).  Accordingly, total excision might not affect the results of this study.  If this was a limitation, it may be an inevitable one with this retrospective study design, because breast tumors with benign CNB results were usually managed with follow-up in clinical practice.  Finally, SWE and color Doppler US in this study were performed by 3 radiologists; and inter-observer variability may exist.

Youk and associates (2017) noted that SWE is a recently developed US technique that can visualize and measure tissue elasticity.  In breast ultrasonography, SWE has been shown to be useful for differentiating benign breast lesions from malignant breast lesions, and it has been suggested that SWE enhances the diagnostic performance of ultrasonography, potentially improving the specificity of conventional ultrasonography using the Breast Imaging Reporting and Data System criteria.  These investigators stated that not only has SWE been shown to be useful for the diagnosis of breast cancer, but has also been shown to provide valuable information that can be used as a pre-operative predictor of the prognosis or response to chemotherapy.  Moreover, the authors stated that although quantitative elasticity information obtained by SWE in addition to B-mode US has improved diagnostic performance, false results have been reported in 6.4 % to 36.6 % of cases, in which the imaging results did not correspond to the pathologic results.  Specifically, the false-positive (FP) rates of benign masses (53 % using qualitative analysis and 22 % to 37 % using quantitative analysis) were reported to be higher than the false-negative (FN) rates of malignant masses (8 % using qualitative analysis and 6 % to 10 % using quantitative analysis).  Considering that benign breast lesions showing FP SWE findings were significantly larger, the FP results can be explained by the size of the breast mass, as larger masses are likely to cause the probe to be unevenly applied to the skin above the masses, which could hinder adequate image acquisition.  Other lesion-related factors contributing to false diagnoses in SWE may include the presence of pure in-situ disease, smaller malignant masses, malignant masses with a circumscribed margin and an abrupt lesion boundary, and the grade of invasive disease.  Aside from the intrinsic tumor characteristics, patient-related or clinical factors associated with false elastography features include the mode of detection (symptomatic versus mammography screening), age at diagnosis, breast thickness, lesion depth, distance from the nipple to the lesion, and the quality of the image.  The authors cautioned that when SWE examinations are performed and interpreted for breast masses, investigators should take into consideration the clinical and lesion-related factors that are associated with inaccurate elastography findings.

Crombe and colleagues (2018) evaluated the ability of SWE to distinguish between benign and malignant palpable masses of the adult male breast.  Clinical examination, mammography, B-mode and Doppler US findings and SWE quantitative parameters were compared in 50 benign lesions (including 40 gynecomastia) and 15 malignant lesions (invasive ductal carcinomas) from 65 patients who were consecutively addressed for specialized advice at the authors’ comprehensive cancer center.  Mean elasticity (El mean), maximum elasticity (El max), El mean of the surrounding fatty tissue and lesion to fat ratio (El ratio) were reported for each patient.  Malignant masses displayed significantly higher El mean (p < 0.0001), El max (p < 0.0001) and El ratio (p < 0.0001) compared to benign masses without overlap of values between the 2 groups.  By adding SWE to clinical examination, mammography and US, all the lesions would have been retrospectively correctly diagnosed as benign or malignant; 1 FP could have been down-staged, 14/65 un-determined masses could have been correctly re-classified as 4 malignant and 10 benign lesions, for which biopsies could have consequently been avoided.  The authors concluded that evaluation of male breast palpable masses by SWE showed that malignant masses were significantly stiffer lesions and may improve diagnostic management when clinical examination, mammography and conventional US were doubtful.  They noted that quantitative SWE is feasible in male breast and could be of great interest to help classify doubtful lesions after classical clinical and radiological evaluations, probably because of different anatomy and different tumors epidemiology compared with female breast.

Balleyguier and associates (2018) evaluated the diagnostic value of MRE in addition to MRI to differentiate malignant from benign breast tumors, and the feasibility of performing MRE on the whole breast.  MRE quantified biomechanical properties within the entire breast (50 slices) using an 11-min acquisition protocol at an isotropic image acquisition resolution of 2 × 2 × 2 mm3.  A total of 50 patients were included; finally, 43 patients (median age of 52 years) with a suspect breast lesion detected by mammography and/or US were examined by MRI and MRE at 1.5 T.  The viscoelastic parameters, i.e., elasticity (Gd), viscosity (Gl), the magnitude of the complex shear modulus Gd2 + Gl2, and the phase angle y = 2πatanGlGd, were measured via MRE and correlated with MRI Breast Imaging-Reporting and Data System (BI-RADS) score, histological type, and histological grade.  Stroma component and angiogenesis were also correlated with viscoelastic properties.  In the 43 lesions, Gd decreased and y increased with the MRI BI-RADS score (pGd  = 0.02, py  = 0.002), whereas (Gl) and y were increased in malignant lesions (pGl  = 0.045, py  = 0.0004).  The AUC increased from 0.84 for MRI BI-RADS alone to 0.92 with the MRI BI-RADS and y (AUC increase +0.08; 95 % CI: -0.003 to 0.16)).  Lesion characterization using the y parameter increased the diagnostic accuracy.  The phase angle y was found to have a significant role (p = 0.01) in predicting malignancy independently of the MRI BI-RADS.  Interestingly, histological analysis showed no correlation between viscoelastic parameters and percentage and type of stroma, CD34 quantification of vessels, or histological grade.  The authors concluded that they have shown for this patient cohort that adding viscoelastic parameters as quantified by MRE to classical MR mammography increased the MRE diagnostic accuracy in lesions left un-determined after mammography and US.  Phase angle was significantly correlated with the malignity of the tumor, and may be a potential functional marker for lesion characterization.  In addition, these researchers had considerably shortened the MRE acquisition time to 10 to 12 mins per patient, while at the same time enabling full breast coverage.  This enabled application of breast MRE in clinical practice.  They stated that studies with larger patient cohorts are needed to further validate these results and to optimize the technique to provide bilateral biomechanical quantification.

The authors stated that this study had several drawbacks.  First, with 43 patients, the sample size was greater than that in some of the previous clinical studies, but still represented a relatively small cohort.  With only 43 patients included in this study and the observed 11 benign lesions, the power was reduced to 25 %.  These researchers had included a wide variety of malignant lesions as well as benign lesions that represent the most frequent lesions generally found in patients.  Nonetheless, the number of patients in each sub‐category was too small, especially invasive lobular carcinoma (ILC), to fully characterize the viscoelastic properties of each lesion type.  Moreover, these investigators did not find any correlation between tumor size and the viscoelastic parameters.  This may be due to the lack of power of the study.  In another previous study, malignant tumors, significantly larger in size than the benign lesions, were quantified as more liquid‐like in MRE.  Additional, MRE was only performed on 1 lesion and on 1 breast in that study.  Current limitations in sequence design and acquisition time prevented data acquisition of the contralateral side.  Additional studies would be needed to develop and optimize the technique.

Jales and colleagues (2018) examined the diagnostic accuracy and clinical consequences of power Doppler morphologic criteria and SWE as complementary imaging methods for evaluation of suspected local breast cancer recurrence in the ipsilateral breast or chest wall.  A total of 32 breast masses with a suspicion of local breast cancer recurrence on B-mode US underwent complementary power Doppler and SWE evaluations.  Power Doppler morphologic criteria were classified as avascular, hypo-vascular, or hyper-vascular; SWE was classified according to a 5-point scale (SWE score) and SWE maximum elasticity.  Diagnostic accuracy was assessed by the sensitivity, specificity, and AUC.  A decision curve analysis assessed clinical consequences of each method.  The reference standard for diagnosis was defined as core needle or excisional biopsy.  Histopathologic examinations revealed 9 (28.2 %) benign and 23 (71.8 %) malignant cases.  Power Doppler US had sensitivity of 34.8 % (95 % CI: 6.6 % to 62.9 %) and specificity of 45.4 % (95 % CI: 19.3 % to 71.5 %).  The SWE score (greater than or equal to 3) had sensitivity of 87.0 % (95 % CI: 66.4 % to 97.2 %) and specificity of 44.4 % (95 % CI: 13.7 % to 78.8 %).  The SWE maximum elasticity (velocity greater than  6.5 cm/s) had sensitivity of 87 % (95 % CI: 66.4 % to 97.2 %) and specificity of 77.8 % (95 % CI: 40.0 % to 97.2 %).  The AUC for the SWE score and SWE maximum elasticity were 0.71 (95 % CI: 0.53 to 0.87) and 0.82 (95 % CI: 0.64 to 0.93), respectively (p = 0.32).  The authors concluded that power Doppler US was unsuitable to discriminate between local breast cancer recurrence and fibrosis.  They stated that although the SWE score and SWE maximum elasticity could make this discrimination, the use of these methods to determine biopsy may lead to poorer clinical outcomes than the current practice of performing biopsies of all suspicious masses.

Furthermore, UpToDate reviews on “Breast density and screening for breast cancer” (Freer and Slanetz, 2018) and “Microinvasive breast carcinoma” (Collins et al,, 2018) do not mention elastography as a management tool.

Dong and co-workers (2019) examined the value of screening breast masses by separate or combined use of multi-modal sound touch elastography (STE).  Women with 159 masses (mean size of 14.86 ± 6.57 mm; range of 5.30 to 30.00 mm) were enrolled in the study.  The pathology results were adopted as diagnostic standards.  The abilities of Young's modulus (E), shear modulus (G), and shear wave (C) to differentiate malignant and benign breast masses based on ROC curves were evaluated, and the optimal cut-off values were obtained.  Sensitivity, specificity, positive likelihood ratio (PLR), and negative likelihood ratio (NLR) were calculated.  Then, the values were combined to perform an overall analysis of STE using evidence-based medicine, construct forest plots, and calculate areas under the summary ROC curves, pooled sensitivity, specificity, PLR, NLR, and diagnostic scores.  A total of 159 masses with a mean size of 14.86 ± 6.57 mm (range of 5.30 to 30.00 mm) were included.  For the various parameters, the diagnostic values were as follows: Gmax > Emax > Cmax > Csd > Esd > Gsd > Emean > Gmean > Cmean .  There were no significant differences in Emin, Gmin, or Cmin.  When the 9 parameters were combined, the pooled sensitivity, specificity, PLR, NLR, diagnostic scores and areas under the summary ROC curves were 84 % (95 % CI: 79 % to 88 %), 82 % (95 % CI: 80 % to 84 %), 4.75 (95% CI, 4.15-5.43), 0.20 (95 % CI: 0.15 to 0.25), 3.19 (95 % CI: 2.84 to 3.54) and 90.2 % (95 % CI: 87 % to 92 %), respectively.  The authors concluded that STE could be recognized as a new US-based diagnostic method for differentiation between benign and malignant breast masses.

Wu and associates (2019) compared the diagnostic value of E and G in the differential diagnosis of benign and malignant breast masses using STE and examined the relationship between G and E in breast lesions.  A total of 96 consecutive women with 110 pathologically confirmed breast masses were included.  All masses were detected by conventional and STE US.  Emean, Emax, Emin, ESD, Gmean, Gmax, Gmin, and GSD were determined and evaluated for evidence of significant differences between benign and malignant breast masses; ROC curves were used to compare the diagnostic efficacy of E and G and to determine the G cut-off value that would help in the differential diagnosis of breast cancer.  Emean, Emax, ESD, Gmean, Gmax, and GSD in cases of malignant breast masses were significantly higher than those in cases of benign masses (p < 0.05).  There was no significant difference between Emin and Gmin (p = 0.565).  In applying the Emean, Emax, ESD, Gmean, Gmax, and GSD to the ROC: the AUC of Gmean and Gmax was greater than the AUC of Emean and Emax, and the AUC of ESD was equal to the AUC of GSD.  The sensitivity and specificity were highest when the Gmean was 10.14 kPa.  They were 84.1 % and 80.3 %, respectively.  The sensitivity and specificity were highest when the Gmax was 52.20 kPa.  They were 88.6% and 87.9%, respectively.  The authors concluded that these preliminary results of STE evaluation of breast masses suggested that the diagnostic value of G is greater than E.

Seo and colleagues (2019) compared the diagnostic performance of B-mode US, SWE, and combined B-mode US and SWE in small breast lesions (less than or equal to 2 cm), and evaluated the factors associated with false SWE results.  A total of 428 small breast lesions (less than or equal to 2 cm) of 415 consecutive patients between August 2013 and February 2017 were included.  The diagnostic performance of each set was evaluated using the AUC analysis.  Histologic diagnosis was used as reference standard.  Multi-variate logistic regression analyses identified the factors associated with false SWE results.  Of 428 lesions, 142 (33.2 %) were malignant and 286 (66.8 %) were benign.  The AUC of the combined modality was higher than that of B-mode US (0.792 versus 0.572, p < 0.001) and that of SWE was higher than that of B-mode US (0.718 versus 0.572, p < 0.001).  Multi-variate analysis showed that the smaller lesion size and in-situ cancer were associated with false negative, and patient's age, high-risk lesion, shorter distance from the skin or chest wall, and deeper breast thickness were associated with false positive (all p < 0.05).  The authors concluded that the addition of SWE to B-mode US could improve the diagnostic performance in less than or equal to 2 cm lesions.  However, US lesion size, pathology, and lesion location were likely to affect the SWE value and result in false results.  The authors noted that despite the diagnostic usefulness of SWE in small breast lesions (less than or equal to 2 cm), US lesion size, pathology, and lesion location were associated with false results.

Au and associates (2019) determined the value of SWE added to targeted US after breast MRI.  From July 2015 to October 2017, a total of 40 patients who underwent targeted US evaluations of suspicious MRI-detected American College of Radiology Breast Imaging Reporting and Data System category 4 lesions (mass or non-mass enhancement) were enrolled in this prospective study.  B-mode US and SWE examinations were performed to detect US correlates to MRI-detected lesions; their Breast Imaging Reporting and Data System categories were recorded; lesions that were dark blue on a 6-point color scale or had maximum elasticity of 30 kPa or less were categorized as soft.  Biopsy was performed with US or MRI guidance, with the pathologic findings correlated with MRI, US, and SWE findings.  The value of SWE for lesion detection and identification of benign lesions was determined.  The mean age of the 40 patients was 51.1 years.  There were 48 MRI-detected lesions (20 cancers, 3 high-risk lesions, and 25 benign lesions); US correlates (8 category 3 and 25 category 4) were shown for 33 lesions (69 %; p < 0.0001), with 16 cancers (80 %; p < 0.0001) and 17 benign lesions; SWE assisted detection of 3 (19 %) cancers on US imaging.  All 7 soft US category 3 lesions were benign (7 of 33 [21 %]; p = 0.0014).  The authors concluded that SWE was useful with targeted US after breast MRI to increase cancer detection by US.  A significant number of US correlates to MRI-detected lesions could have been identified as benign (category 3 and soft) before biopsy, with the potential of short-interval follow-up of MRI-detected lesions with benign US correlates instead of biopsy.

Filip and co-workers (2019) evaluated strain elastography as a complementary tool for characterization of lesions identified during 2nd-look MRI-directed sonographic examination.  These investigators reviewed 83 breast lesions evaluated with MRI, 2nd-look US and strain elastography in 75 consecutive patients (median age of 56 years); US-guided biopsies were performed in all cases.  After histopathological examination, 44 lesions were benign, 38 were malignant and 1 was high-risk.  At MRI, the mean size of the lesions was 12 mm; 40 lesions (48.2 %) appeared as masses, 30 (36.1 %) as "non-masses" and 13 (15.7 %) as "foci".  At 2nd-look US examination, 56 (67.5 %) appeared as masses (mean size of 7 mm) and 27 (32.5 %) as non-masses (mean size of 14 mm).  At strain elastography, among the 39 malignant/high risk lesions, 5 (12.8 %) had a score of 4 or 5, whereas 16 (41 %) had a score of 1 and 2 (false-negative).  Among the 44 benign lesions, 36 (82 %) had a score of 1 or 2, whereas none had a score of 5.  Sensitivity and specificity of strain elastography in the diagnosis of breast cancer were 58 % and 81 %, respectively.  The authors concluded that addition of strain elastography offered no benefit in the characterization of lesions identified on 2nd-look US after breast MRI.

Forte and colleagues (2020) stated that breast cancer-related lymphedema (BCRL) incidence has been increasing over time.  At the present time, there is no preferred imaging tool for diagnosis, staging, and evaluation of the disease.  These researchers examined the use of ultrasound elastography (UE) in BCRL patients.  They carried out a systematic review by querying PubMed, Embase, Ovid Healthstar, and Ovid Medline databases for studies that examined the use of UE in BCRL.  The keywords "elastography" AND "lymphedema" in titles and abstracts were used for the search.  The search retrieved 12, 12, 5 and 6 articles in each database, respectively.  From these, only 4 met the inclusion criteria -- UE methods included 2D strain imaging, SWE, and global UE.  Two of the studies examined the use of UE in the evaluation of BCRL, while only 1 considered its use for diagnosis and staging.  The authors concluded that UE appeared to be a great tool in the evaluation of BCRL to differentiate affected from non-affected arms.  Moreover, these researchers stated that further studies with greater sample sizes are needed to confirm the use of UE in staging and diagnosis of early BCRL stages.

The authors stated that they summarized the findings of different methods of UE; thus, the limitation of heterogeneity and possible biases proper of each study that they found via the risk of bias assessment should be considered.  Another inherent drawback of the review was the potential for search, selection, and publication biases; however, this systematic review was entirely descriptive, in alignment with the purpose of the study.  To the authors’ knowledge, this was the 1st systematic review examining the use of UE for BCRL.

Huang and associates (2020) noted that studies have reported inconsistent findings regarding the diagnostic performance of UE for axillary lymph node metastasis (ALNM) in breast cancer.  In a meta-analysis, these researchers examined the diagnostic performance of UE (divided into quantitative and qualitative elastography) for ALNM in patients with breast cancer.  The PubMed and Embase databases were searched for eligible studies exploring the diagnostic performance of UE for ALNM in patients with breast cancer.  The included studies were divided into quantitative and qualitative elastography groups to perform separate meta-analyses.  The diagnostic performance was examined with pooled sensitivity and specificity and diagnostic odds ratio (DOR) using a bi-variate mixed-effects regression model.  A summary receiver operating characteristic curve was constructed, and the area under the curve (AUC) was calculated.  A total of 7 and 11 studies were included in the quantitative and qualitative elastography meta-analyses, respectively.  The pooled sensitivity and specificity, DOR, and AUC with their corresponding 95 % CI were 0.82 (0.75 to 0.87), 0.88 (0.78 to 0.93), 33 (13 to 83), and 0.89 (0.86 to 0.91), respectively, for quantitative elastography and 0.81 (0.69 to 0.89), 0.92 (0.79 to 0.97), 46 (12 to 181), and 0.92 (0.89 to 0.94), respectively, for qualitative elastography.  No significant publication bias existed; and Fagan plots demonstrated good clinical utility; however, substantial heterogeneity existed among studies.  Study design, measurement, and reference standard served as potential sources of heterogeneity for quantitative studies, which were measurement and reference standard for qualitative studies.  The authors concluded that both quantitative and qualitative elastography appeared to be feasible, non-invasive diagnostic tools for ALNM in breast cancer; however, because of the substantial heterogeneity among these studies, evidence of data reliability is still insufficient.  These researchers stated that these findings must be interpreted carefully, paying particular attention to heterogeneity issues, especially for quantitative elastography studies.  They stated that adequate method for higher accuracy of ALNM is needed in the future; and prospective, multi-center population-based trials are needed to confirm the diagnostic value of UE for ALNM in breast cancer.

The authors stated that this study had several drawbacks.  First, studies published in languages other than English were not included in this meta-analysis; however, this did not appear to produce significant publication bias in either the quantitative or the qualitative elastography studies.  Second, substantial heterogeneity was observed among both the quantitative and qualitative elastography studies, which would limit recommending their integration into clinical practice.  Theoretically, the funnel plot was invalid because the accuracy was low when fewer than 10 studies were available for the quantitative summary; therefore, the number of quantitative studies (n = 7) was too small to conduct statistical analyses such as publication bias and analysis for the heterogeneity.  In addition, according to the results of quality assessment, 2 studies with high risk of bias were low in quality, which possibly reduced the accuracy of bias assessment.  Third, the US machines of the included studies were not exactly the same.  Technological improvements or system errors may have led to measurement bias.  Furthermore, the evaluation and diagnostic performance of UE rely on the experience and skills of the operators as well as the operating procedures.  All elastography imaging techniques have a learning curve for implementation to daily practice; thus, especially for new operators, the diagnostic performance of these imaging tools may not be satisfactory as concluded in this meta-analysis.  All these drawbacks of UE serve as potential sources of heterogeneity.

Jung and co-workers (2021) noted that capsular contractures around breast implants usually develop leading to pain and aesthetically inadequate results and often requires the replacement of the implants.  Textured silicone implants are the most commonly placed implant, but polyurethane-coated implants are increasingly being used in an attempt to ameliorate the long-term complications associated with implant insertion.  Capsular contracture is traditionally classified using the Baker scale, a subjective classification system based upon clinical findings.  These researchers examined the association between pain due capsular contraction, Baker Score and different techniques of UE.  Patients were contacted who had undergone an implant replacement due to capsular contracture.  Inclusion criterion was the re-implantation of a PU-coated implant.  In the 3rd year after changing the implant a follow-up examination was carried out in 16 patients with 23 implants.  A conventional examination with anamnesis, tactile and visual findings to obtain a Baker score, and US examinations including SWE, acoustic radiation force impulse (ARFI) and compound elastography were performed.  Furthermore, pain was evaluated using a visual analogue scale (VAS).  The pain data showed a significant improvement (before implant exchange: 4.1 ± 2.8 score points) with significance in favor of the current state (1.7 ± 1.0 pain score points; p = 0.002).  All patients suffered from less or no pain 3 years after exchange of the implant.  Pain values and elastography (ARFI values) correlated well (r = 0.873), with increasing Baker score the ARFI values increased.  UE evaluations could locally determine tissue density but correlated only to a limited extent with the test findings according to Baker; UE values of mammary gland tissue without implant did not differ from mammary gland tissue around implants.  The authors concluded that pre-operative Baker scores prior to exchange and the current Baker scores at the follow-up showed significantly lower score points 3 years after exchange of the implants; UE appeared to be an objective classification of capsular fibrosis.  Moreover, these researchers stated that these first results motivated them to initiate a prospective, multi-center trial.

Wu et al (2022) stated that SWE and strain elastography (SE) are 2 new US technologies that have developed rapidly in recent years.  Elastography transforms the elastic information of tissue into optical information for display; therefore, providing more intuitive display of tissue elasticity.  Conflicting results have been obtained in different researchers' reports on the accuracy comparison of the 2 elastography technologies in the diagnosis of breast tumors.  In a meta-analysis, these investigators compared the accuracy of the 2 elastography technologies in the diagnosis of breast tumors; and provided a reference for clinical decision-making.  They searched Chinese and English studies on the accuracy of SWE and SE in the diagnosis of breast tumors from PubMed, Web of Science, China national knowledge infrastructure and Wanfang databases, and the time was up to December 30, 2020.  Two literature reviewers screened the studies according to the screening criteria, and Quality Assessment of Diagnostic Accuracy Study tool was used to evaluate the quality of included literatures.  Meta Disc1.4 and Stata14.0 software were used to conduct heterogeneity test, sensitivity analysis and publication bias test.  A total of 10 studies (1,599 patients and 1,709 breast lesions) were included in this analysis.  The final results in the SWE were as follow: The pooled sensitivity (Sen) was 0.852 (95 % CI: 0.826 to 0.874), the pooled specificity (Spe) was 0.799 (95 % CI: 0.776 to 0.820), the pooled PLR was 4.758 (95 % CI: 3.443 to 6.576), the pooled NLR was 0.192 (95 % CI: 0.147 to 0.250), the pooled DOR was 29.071 (95 % CI: 16.967 to 49.811), and the AUC-ROC was 0.9159.  The final results in the SE were as follow: The pooled Sen was 0.843 (95 % CI: 0.817 to 0.866), the pooled Spe was 0.766 (95 % CI: 0.743 to 0.789), the pooled PLR was 4.387 (95 % CI: 3.088 to 6.233), the pooled NLR was 0.216 (95 % CI: 0.179 to 0.261), the pooled DOR was 22.610 (95 % CI: 15.622 to 32.724), and the AUC-ROC was 0.8987.  The authors concluded that the Sen and Spe of SWE were higher than those of SE, suggesting that SWE may have a higher accuracy in the diagnosis of breast tumors; , and the use of SWE combined with the conventional US could further enhance the accuracy in diagnosing breast tumors.

The authors stated that this analysis had 2 main drawbacks.  First, because the related research was less, the sample size differences of included studies resulted in heterogeneity, and the quality of the included studies was low; thus, this review lacked sufficient statistical support to examine the results of the accuracy of the 2 types of elasticity imaging diagnosis of breast tumor.  Second, meta-analysis is a retrospective study that may lead to subject selection bias.  Retrieving Chinese-English studies will lead to bias in research selection.  Notably, most of the included studies was from Asia, which could adversely affect the reliability and validity of these findings. 

Chen et al (2022) noted that several studies have reported the prognostic value of ultrasound elastography (UE) in patients receiving neoadjuvant chemotherapy (NACT) for breast cancer.  However, the assessment of parameters differed between SWE and SE in terms of measured elasticity parameter and mode of imaging.  Thus, it is important to examine the accuracy of the 2 modes of elastography.  In a systematic review and meta-analysis, these researchers examined the accuracy of UE for predicting the pathologic complete response (pCR) in breast cancer patients following NACT.  They carried out a comprehensive and systematic search in the databases of Medline, Embase, SCOPUS, PubMed Central, CINAHL, Web of Science and Cochrane library from inception until December 2020.  Meta-analysis was carried out using STATA software "Midas" package.  A total of 14 studies with 989 patients were included.  The pooled Sen were 86 % (95 % CI: 76 % to 92%) for UE, 77 % (95 % CI: 68 % to 84 %) for SWE, and 92 % (95 % CI: 73 % to 98 %) for SE.  The pooled score Spe were 86 % (95 % CI: 80 % to 90%) for UE, 84 % (95 % CI: 72 % to 91 %) for SWE, and 87 % (95 % CI: 81 % to 92 %) for SE.  A significant heterogeneity was found among studies based on the Chi-square test results and an I2 statistic of greater than 75 %.  The authors concluded that strain-wave type of UE could accurately predict the pCR following NACT amongst breast cancer patients.  Moreover, these researchers stated that studies examining its accuracy in different ethnic populations are needed to strengthen the evidence.  Furthermore, these investigators stated that future studies perhaps should attempt to combine palpation imaging, US, and mammography to analyze ambiguous clinical cases in order to improve breast lesion diagnosis; and additional large-scale setting-specific longitudinal studies are needed to establish the best imaging methods to examine all the patients administered with NACT.

The authors stated that this meta-analysis had several drawbacks.  First, there was a significant between-study variability in this analysis, which could limit the prospect to infer or interpret the pooled findings.  However, these investigators examined the source of heterogeneity using meta-regression analysis to overcome this limitation.  Second, the predictive accuracy of UE depends on several other factors such as the ethnicity, timing of the index test and outcome assessment, and disease severity.  However, these investigators could not examine their influence in this analysis.  Third, the number of subjects/participants included was relatively small.

In a systematic review and meta-analysis, Pillai et al (2022) examined the diagnostic accuracy of 2D-SWE for differentiating benign and malignant breast lesions in women with abnormal findings on mammography.  These investigators included studies of diagnostic accuracy published before June 2021 using 2D-SWE to examine female breast lesions.  Included studies were required to include at least 50 lesions, report quantitative shear-wave speed (SWS) thresholds, and include a reference standard of either biopsy or 2-year stability.  Included studies used the mean, maximum, minimum, or SD of SWS for classification.  A systematic search of PubMed, Scopus, Embase, Ovid-Medline, the Cochrane Library, and Web of Science was performed.  Bias and applicability of the studies were assessed using Quality Assessment of Diagnostic Accuracy Studies 2.  A hierarchical summary ROC model was used to arrive at the summary statistics.  A total of 87 prospective and retrospective studies were included, encompassing 17,810 women (mean age of 42.3 ± 10.4 years) with 19,043 lesions (7,623 malignant).  Summary Sen and Spe, respectively, were 0.86 (95 % CI: 0.83 to 0.88) and 0.87 (95 % CI: 0.84 to 0.88) for mean SWS, 0.83 (95 % CI: 0.80 to 0.85) and 0.88 (95 % CI: 0.86 to 0.90) for the maximum, 0.86 (95 % CI: 0.74 to 0.93) and 0.81 (95 % CI: 0.69 to 0.89) for the minimum, and 0.82 (95 % CI: 0.77 to 0.86) and 0.88 (95 % CI: 0.85 to 0.91) for the SD.  Alternatively, the AUC-ROC were 0.93 (95 % CI: 0.91 to 0.94), 0.92 (95 % CI: 0.90 to 0.94), 0.90 (95 % CI: 0.82 to 0.96), and 0.92 (95 % CI: 0.88 to 0.94), respectively.  The authors concluded that this review demonstrated the discriminative power of SWE in the diagnosis of breast cancer.  Using the resulting likelihood ratios, SWE may prove beneficial in down-grading BI-RADS 4a or up-grading BI-RADS 3 lesions. However, current society guidelines do not provide definitive recommendations regarding the use of SWE and its counterpart SE.  These researchers stated that comparison with these findings suggested that SE alone or a combination of SE and SWE may provide better diagnostic performance than SWE alone and serve as an adjunct to current diagnostic techniques.

Shahzad et al (2022) noted that conventional B-mode breast US, though the primary modality to determine benign or malignant nature of a solid breast lesion, sometimes encounters overlapping sonographic morphological features in a single lesion.  Ina single-center study, these researchers examined the role of SE and SWE in differentiating benign and malignant breast lesions.  Patients meeting inclusion criteria referred to the authors’ hospital for US followed by biopsy or surgical excisions were examined with B-mode US and by both SE and SWE.  Main outcome measures included mean values of SE and SWE in benign and malignant breast lesions, determination of cut-off using AUC curves as well as Sen and Spe of both techniques. A total of 100 breast lesions from 95 consecutive patients were included in this study.  The mean (SD) SE ratio in the overall patient population was 4.1 (2.0).  Cut-off for benign versus malignant lesions was 2.86 on the ROC curve.  The AUC was 0.911 (95 % CI: 0.835 to 0.988: SE, 0.039) with a Sen of 95.8 % and a Spe of 89.3 %.  For the SWE kPa values, the ROC curve showed the AUC was 0.929 (95 % CI: 0.870 to 0.988; SE: 0.030, p < 0.001).  Assigning 45.3 as a cut-off value provided a Sen of 95.8 % with a Spe of 85.7 %; the PPV was 94.5 % and the NPV was 89.6 %.  The Breast Imaging Reporting and Data System (BI-RADS) category alone was able to differentiate between benign and malignant lesions with a Sen of 91.7 % and a Spe of 100 % keeping the cut-off value between 4a and 4b.  The area under the ROC curve was 0.979.  Combining the 3 (BI-RADS + SE + SWE) distinguished benign versus malignant lesions with a Sen up to 100 % and Spe up to 96.3 %.  The authors concluded that combining SE and SWE as a complementary tool with conventional B-mode US has a significant potential for better characterization of solid breast lesions and decreasing unnecessary biopsies of BI-RADS IVa lesions.  The authors stated that the main drawback of this study was that it was a s single-center study.

In a meta-analysis, Wang and Wang (2023) examined the value of UE in the differential diagnosis of the triple-negative breast cancer (TNBC) and non-TNBC.  These investigators searched the PubMed, Cochrane Library, and CBM databases from inception to July 20, 2022 and used STATA version 14.0 and Meta-Disc version 1.4 software.  They computed summary statistics for Sen, Spe, PLR and NLR (LR+/LR-), DOR, and summary ROC curves.  Cochran Q-statistic and I2 test were used to evaluate potential heterogeneity between studies.  Sen analysis was performed to examine the effect of a single study on overall estimation.  These researchers also carried out a meta regression analysis to examine potential sources of heterogeneity.  A total of 9 studies that met all the criteria for acceptance were incorporated into the meta-analysis.  TNBC (n = 317) and non-TNBC (n = 1,055) were evaluated.  All breast tumors were histologically confirmed.  The pooled Sen was 0.78 (95 % CI: 0.58 to 0.90); the pooled Spe was 0.86 (95 % CI: 0.78 to 0.91).  The pooled LR+ was 5.46 (95 % CI: 3.07 to 9.73); the pooled LR- was 0.26 (95 % CI: 0.12 to 0.55).  The pooled DOR of UE was 21.00 (95 % CI: 6.14 to 71.78).  The AUC-ROC curve was 0.89 (SE = 0.0378).  No evidence was found to reveal bias (t = 0.10, p = 0.92).  The authors concluded that the findings of this meta-analysis showed that UE could have high diagnostic accuracy in distinguishing TNBC and non-TNBC.  Moreover, these researchers stated that further detailed studies are needed to confirm these findings.

The authors stated that this study had several drawbacks.  First, owing to the relatively small sample sizes and low quality of the included studies, there was insufficient data to evaluate the accuracy of UE.  Second, the inclusion of studies with histological confirmation only and the retrospective nature of the meta-analysis could have led to subject selection bias.  Third, the majority of the included studies originated from China, which may adversely affect the reliability and validity of these findings.

The authors stated that this study had several drawbacks.  First, retrospective study designs have inherent limitations, such as selection bias and recall bias.  Second, their case load was too small.  Third, shell functionality was limited to 1 instrument, which was not conducive to the generalization of the results.

Guo et al (2023) noted that several studies have reported the difficulties in distinguishing malignant lesions of the breast from benign lesions due to overlapping morphological features on US.  In a prospective, single-center study, these researchers developed a nomogram based on (SWE, Angio Planewave Ultrasensitive imaging (Angio PLUS (AP)), and conventional US imaging biomarkers to predict malignancy in patients with breast lesions. This trial included 117 female patients with suspicious lesions of the breast.  Features of lesions were extracted from SWE, AP, and conventional US images.  The least absolute shrinkage and selection operator (Lasso) algorithms were used to select breast cancer-related imaging biomarkers, and a nomogram was developed based on 6 of the 16 imaging biomarkers.  This model exhibited good discrimination (AUC: 0.969; 95 % CI: 0.928 to 0.989) between malignant and benign breast lesions.  Moreover, the nomogram also demonstrated good calibration and clinical usefulness.  The authors concluded that their nomogram could be a potentially useful tool for individually-tailored diagnosis of breast tumors in clinical practice.  Moreover, these researchers stated that future large-scale, multi-center studies are needed to add to the database.

The authors stated that this study had several drawbacks.  First, this trial was a prospective, single-center cohort study with a relatively small sample size (n = 117), which limited the generalizability of the findings.  Second, there were no external validation datasets available, although these investigators carried out the 500 boot-strap samples for internal validation.  Hence, this prediction tool appeared to have reasonable clinical utility; however, further validation in external datasets is needed.  Third, limited by the size of the sampling frame of the SWE image, these researchers only included breast masses with the largest diameter of 30 mm or more, which may have introduced a selection bias.  They stated that further comprehensive studies are needed to resolve these issues.

Optical Coherence Micro-Elastography for Intraoperative Assessment of Breast Cancer Margins and Lymph Node Involvement

Allen and colleagues (2016) noted that incomplete excision of malignant tissue is a major issue in breast-conserving surgery, with typically 20 to 30 % of cases requiring a second surgical procedure arising from post-operative detection of an involved margin. These investigators reported advances in the development of a new intra-operative tool, optical coherence micro-elastography (OCME), for the assessment of tumor margins on the micro-scale.  They demonstrated an important step by conducting whole specimen imaging in intra-operative time frames with a wide-field scanning system acquiring mosaicked elastograms with overall dimensions of approximately 50 × 50 mm, large enough to image an entire face of most lumpectomy specimens.  This capability was enabled by a wide-aperture annular actuator with an internal diameter of 65 mm.  The authors demonstrated feasibility by presenting elastograms recorded from freshly excised human breast tissue, including from a mastectomy, lumpectomies and a cavity shaving.  These researchers stated that they will incorporate further reductions in acquisition and processing time and determine the diagnostic accuracy of compression OCME in identifying malignant tissue during breast-conserving surgery.

Kennedy and associates (2016) noted that evaluation of lymph node involvement is an important factor in detecting metastasis and deciding whether to perform axillary lymph node dissection (ALND) in breast cancer surgery.  As ALND is associated with potentially severe long-term morbidity, the accuracy of lymph node assessment is imperative in avoiding unnecessary ALND.  The mechanical properties of malignant lymph nodes are often distinct from those of normal nodes.  A method to image the micro-scale mechanical properties of lymph nodes could, therefore, provide diagnostic information to assist in the evaluation of lymph node involvement in metastatic cancer.  In this study, these investigators scanned axillary lymph nodes, freshly excised from breast cancer patients, with OCME, a method of imaging micro-scale mechanical strain, to assess its potential for the intra-operative assessment of lymph node involvement.  A total of 26 fresh, unstained lymph nodes were imaged from 15 patients undergoing mastectomy or breast-conserving surgery with axillary clearance.  Lymph node specimens were bisected to allow imaging of the internal face of each node.  Co-located OCME and optical coherence tomography (OCT) scans were taken of each sample, and the results compared to standard post-operative hematoxylin-and-eosin-stained histology.  The optical back-scattering signal provided by OCT alone may not provide reliable differentiation by inspection between benign and malignant lymphoid tissue.  Alternatively, OCME highlighted local changes in tissue strain that corresponded to malignancy and were distinct from strain patterns in benign lymphoid tissue.  The mechanical contrast provided by OCME complemented the optical contrast provided by OCT and aided in the differentiation of malignant tumor from uninvolved lymphoid tissue.  The authors concluded that the combination of OCME and OCT images represents a promising method for the identification of malignant lymphoid tissue.  They noted that this method shows potential to provide intra-operative assessment of lymph node involvement, thus, preventing unnecessary removal of uninvolved tissues and improving patient outcomes.

Gubarkova and colleagues (2020) stated that the possibility to evaluate molecular-biological and morphological features of particular breast cancer types can improve the precision of resection margin detection and enable accurate determining of the tumor aggressiveness, which is important for treatment selection.  To enable reliable differentiation of breast-cancer subtypes and evaluation of resection margin, without performing conventional histological procedures, these researchers employed cross-polarization OCT (CP-OCT) and compared it with a novel variant of compressional OCT (C-OCE) in terms of the diagnostic accuracy (Ac) with histological verification.  The study used 70 excised breast cancer specimens with different morphological structure and molecular status (Luminal A, Luminal B, Her2/Neo+, non-luminal and triple-negative cancer).  The primary objective was to formulate convenient criteria of visual assessment of CP-OCT and C-OCE images intended to first differentiate tumorous and non-tumorous tissues; and second to enable more precise differentiation among different malignant states.  These investigators identified such criteria based on the presence of heterogeneities and characteristics of signal attenuation in CP-OCT images, as well as the presence of inclusions/mosaic structures combined with visually feasible assessment of several stiffness grades in C-OCE images.  They also carried out a blinded reader study of the Ac of C-OCE versus CP-OCT, for delineation of tumorous versus non-tumorous tissues followed by identification of breast cancer subtypes.  For tumor detection, C-OCE showed higher specificity than CP-OCT (97.5 % versus 93.3 %) and higher Ac (96.0 % versus 92.4 %).  For the first time, the Ac of C-OCE and CP-OCT were examined for differentiation between non-invasive and invasive breast cancer (90.4 % and 82.5 %, respectively).  In addition, for invasive cancers, the difference between invasive but low-aggressive and highly-aggressive subtypes can be detected.  For differentiation between non-tumorous tissue and low-aggressive breast-cancer subtypes, Ac was 95.7 % for C-OCE and 88.1 % for CP-OCT.  For differentiation between non-tumorous tissue and highly-aggressive breast cancers, Ac was found to be 98.3 % for C-OCE and 97.2 % for CP-OCT.  In all cases C-OCE showed better diagnostic parameters independently of the tumor type.  The authors concluded that these findings confirmed the high potential of OCT-based examinations for rapid and accurate diagnostics during breast conservation surgery.

Combined Shear-Wave Elastography with Ultrasonography for Detection of Breast Cancer

Di Meo and co-workers (2020) showed, for the first time, a triple-mode scan using electromagnetic waves, in the shape of millimeter waves, and US waves, to obtain B-mode and quasi-static elastography images of a phantom of human breast tissues.  A homogeneous phantom composed of non-toxic, low-cost and easy-to-handle materials (i.e., water, oil, gelatin and dishwashing liquid) was produced, with an inclusion made of water and agar.  These were intended to mimic, in terms of dielectric properties, healthy adipose tissues and neoplastic tissues, respectively.  A millimeter-wave imaging prototype was used to scan the phantom, by implementing a linear synthetic array of 24 antennas with a central working frequency of 30-GHz.  The phantom was then scanned using an US research system and a linear-array probe at 7-MHz, acquiring both B-mode and quasi-static elastography images.  The millimeter-wave system showed an excellent ability to detect the target placed at about 1.4 cm depth.  Furthermore, in the US case the inclusion was correctly detected as a hypo-echoic, stiff mass.  This first experimental findings showed that millimeter-wave, US and elasticity imaging could be used jointly to detect tumor-like targets into phantoms mimicking healthy breast tissues.  The authors concluded that they provided promising preliminary findings to further examine the application of this multi-modal approach in all those critical cases in which such complementary imaging techniques could be exploited for an enhanced tumor detection, based on tissues dielectric, acoustic and elastic properties.

In a systematic review and meta-analysis, Park and colleagues (2021) compared the diagnostic performance and biopsy reduction rate of combined SWE and B-mode US versus B-mode US alone for detection of breast lesions and determined the most discriminatory parameter in SWE.  The resources for the study were obtained from Medline, Embase, Cochrane Library, and KoreaMed on August 17, 2018.  The quality of the articles was examined using the Scottish Intercollegiate Guidelines Network (SIGN) tool.  A total of 25 articles with 5,147 breast lesions were selected.  The meta-analysis showed pooled sensitivities of 0.94 and 0.97 (p = 0.083), pooled specificities of 0.85 and 0.61 (p = 0.009), and AUC of 0.96 and 0.96 (p = 0.095) for combined SWE and B-mode US versus B-mode US alone.  When SWE was combined with B-mode US, the Breast Imaging Reporting and Data System category changed from 4 to 3 in 71.3 % of the tests, decreasing the frequency of unnecessary biopsies by 41.1 %.  All 4 parameters of SWE (the color grade of lesion stiffness, maximum elasticity, mean elasticity, and color grade of lesion stiffness/homogeneity of the lesion) improved the specificity when they were added to B-mode US.  The AUC for each SWE parameter was 0.99, 0.96, 0.96, and 0.93, respectively.  The authors concluded that adding SWE to B-mode US not only provided additional diagnostic information for differentiating between benign and malignant breast lesions, but also decreased the likelihood of unnecessary biopsies.  Moreover, these researchers stated that future studies are needed to extend these findings and to confirm appropriate cut-off values of SWE.

The authors stated that this study had several drawbacks.  First, the cut-off values for SWE ranged from 30 to 145.9 kPa across the different studies.  These investigators were unable to obtain an optimal single cut-off value because of heterogeneity among the studies; therefore, this issue remained to be resolved via studies with a larger sample size.  Second, although these researchers carried out a validated meta-analysis, considerable heterogeneity among studies was observed.  Nonetheless, they conducted a comprehensive systematic review with robust methods, and no publication bias was identified among the selected studies.  Third, these researchers did not examine the performance of SWE in up-grading lesions; thus, any future studies need to consider the various factors that could affect diagnostic accuracy.  Furthermore, different techniques such as ARFI were not evaluated in this meta-analysis.

Toprak and associates (2021) noted that idiopathic granulomatous mastitis (IGM) is a chronic, unpleasant autoimmune inflammatory condition and is clinically and radiologically often confused with breast malignancy.  These investigators examined the contributions of qualitative and quantitative aspects of ARFI elastography to the differential diagnosis between IGM and invasive ductal carcinoma (IDC) in the breast.  A total of 94 women with IDC and 39 with IGM were included in the study.  Shear wave velocity (SWV) was calculated for all lesions using quantitative elastography.  Next, each lesion's correspondence on qualitative elastographic images to those on the B-mode US images was evaluated: pattern 1, no findings on elastography images; pattern 2, lesions that were bright inside; pattern 3, lesions that contained both bright and dark areas; and pattern 4, lesions that were dark inside.  Pattern 4 was sub-divided into 4a (dark area same size as lesion) and 4b (dark area larger than lesion size).  Sensitivity, specificity, PPV, NPV, and diagnostic accuracy were calculated.  The mean SWV based on ARFI elastography was 3.78 ± 1.26 m/s for IGM and 5.34 ± 1.43 m/s for IDC lesions (p < 0.05).  Based on qualitative ARFI elastography, IDC lesions were mostly classified as pattern 4b, while IGM lesions were mostly classified as pattern 1 or 2 (p = 0.01).  Evaluation of both the qualitative and quantitative aspects of ARFI elastography yielded a sensitivity of 89 % and specificity of 84 %.  The authors concluded that ARFI elastography may facilitate the differential diagnosis between IGM and IDC.

Non-Elastographic Quantitative Ultrasound Tissue Characterization for Screening and Diagnosis of Breast Cancer

Byra and colleagues (2016) noted that statistical modeling of an ultrasound (US) back-scattered echo envelope is used for tissue characterization.  However, in the presence of complex structures within the analyzed area, estimation of parameters is disturbed and unreliable (e.g., in the case of breast tumor classification).  In order to improve the differentiation of breast lesions, these researchers proposed a method based on the segmentation of homodyned K distribution parameter maps.  Regions within lesions of different scattering properties were extracted and analyzed.  In order to improve the classification, the best-performing features were selected from various regions and then combined.  A radiofrequency (RF) dataset consisting of 103 breast lesions was used in this analysis.  Maps of homodyned K distribution parameters were created using an algorithm based on signal-to-noise ratio, kurtosis, and skewness of fractional-order envelope moments.  A Markov random field model was used to segment parametric maps.  Features of different segments were extracted and examined based on bootstrapping and the ROC curve.  To determine the best-performing feature subset, these investigators employed the joint mutual information criterion.  It was found that there were individual features that performed better than the ones commonly used for lesion characterization, like the parameter obtained via averaging of values over the whole lesion.  These researchers selected and discussed the best-performing features.  Properties of different extracted regions were important and improved the distinction between benign and malignant tumors.  The best performance was obtained by combining 4 features with the area under the ROC curve (AUC) of 0.84.  The authors concluded that the findings of this study showed that the analysis of internal changes in lesion parametric maps led to a better classification of breast tumors.  They recommended combining multiple features for characterization, instead of using only 1 parameter, especially in the case of heterogeneous lesions.

Osapoetra and co-workers (2020) stated that accurate and timely diagnosis of breast cancer is important because of its high incidence and high morbidity.  Screening can improve overall prognosis by detecting the disease early.  Biopsy remains as the gold standard for pathological confirmation of malignancy and tumor grading.  The development of diagnostic imaging techniques as an alternative for the rapid and accurate characterization of breast masses is needed.  Quantitative US (QUS) spectroscopy is a modality well suited for this purpose.  These researchers examined different texture analysis methods used on QUS spectral parametric images for the characterization of breast lesions.  Parametric images of mid-band-fit (MBF), spectral-slope (SS), spectral-intercept (SI), average scatterer diameter (ASD), and average acoustic concentration (AAC) were determined using QUS spectroscopy from 193 patients with breast lesions.  Texture methods were employed to quantify heterogeneities of the parametric images.  Three statistical-based approaches for texture analysis that include Gray Level Co-occurrence Matrix (GLCM), Gray Level Run-length Matrix (GRLM), and Gray Level Size Zone Matrix (GLSZM) methods were evaluated.  QUS and texture-parameters were determined from both tumor core and a 5-mm tumor margin; and were used in comparison to histopathological analysis in order to classify breast lesions as either benign or malignant.  These investigators developed a diagnostic model using different classification algorithms including linear discriminant analysis (LDA), k-nearest neighbors (KNN), support vector machine with radial basis function kernel (SVM-RBF), and an artificial neural network (ANN).  Model performance was examined using leave-one-out cross-validation (LOOCV) and hold-out validation.  Classifier performances ranged from 73 % to 91 % in terms of accuracy dependent on tumor margin inclusion as well as classifier methodology.  Using information from tumor core alone, the ANN achieved the best classification performance of 93 % sensitivity, 88 % specificity, 91 % accuracy, 0.95 AUC using QUS parameters and their GLSZM texture features.  The authors concluded that a QUS-based framework and texture analysis methods enabled classification of breast lesions with greater than 90 % accuracy.  These findings suggested that different methods for extracting textural features of QUS spectral parametric images could result in the same classification performance to that obtained using more computationally intensive derivative texture methods.

Osapoetra and colleagues (2021) developed a multi-centric response predictive model using QUS spectral parametric imaging and novel texture-derivate methods for determining tumor responses to NAC before commencement of therapy.  QUS spectroscopy provided parametric images of MBF, SS, SI, ASD, and AAC in 78 patients with locally advanced breast cancer (LABC) undergoing NAC.  Ultrasound RF data were collected from Sunnybrook Health Sciences Center (SHSC), University of Texas MD Anderson Cancer Center (MD-ACC), and St. Michaels Hospital (SMH) using 2 different systems.  Texture analysis was used to quantify heterogeneities of QUS parametric images.  Furthermore, a 2nd-pass texture analysis was employed to obtain texture-derivate features.  QUS, texture- and texture-derivate parameters were determined from both tumor core and a 5-mm tumor margin and were used in comparison to histopathological analysis for developing a response predictive model to classify responders versus non-responders.  Model performance was assessed using leave-one-out cross-validation.  Three standard classification algorithms including a LDA, k- KNN, and SVM-RBF were evaluated.  A combination of tumor core and margin classification resulted in a peak response prediction performance of 88 % sensitivity, 78 % specificity, 84 % accuracy, 0.86 AUC, 84 % PPV, and 83 % NPV, achieved using the SVM-RBF classification algorithm.  Other parameters and classifiers performed less well running from 66 % to 80 % accuracy.  The authors concluded that this study demonstrated a working a priori response predictive model (84 % accuracy and 0.86 AUC), developed using QUS and texture-derivate parameters that were determined from US RF data obtained from multiple cancer centers.  Model development using multi-centric data provided an indication towards generalizing the QUS framework.  These researchers stated that this work paves the way towards an implementation of QUS and texture analyses imaging tools for oncologists to provide customized cancer treatment adjusted to specific patient.

Taleghamar and associates (2021) noted that the effectiveness of QUS multi-parametric imaging in conjunction with un-supervised classification algorithms was examined for the 1st time in characterizing intra-tumor regions to predict breast tumor response to chemotherapy prior to the commencement of treatment.  QUS multi-parametric images of breast tumors were generated using the US RF data acquired from 181 patients diagnosed with locally advanced breast cancer and planned for NAC followed by surgery.  A hidden Markov random field (HMRF) expectation maximization (EM) algorithm was used to identify distinct intra-tumor regions on QUS multi-parametric images.  Several features were extracted from the segmented intra-tumor regions and tumor margin on different parametric images.  A multi-step feature selection procedure was employed to construct a QUS biomarker consisting of 4 features for response prediction.  Evaluation results on an independent test set indicated that the developed biomarker coupled with a decision-tree model with adaptive boosting (AdaBoost) as the classifier could predict the treatment response of patient at pre-treatment with an accuracy of 85.4 % and an AUC of 0.89.  In comparison, the biomarkers consisted of the features derived from the entire tumor core (without consideration of the intra-tumor regions), and the entire tumor core and the tumor margin could predict the treatment response of patients with an accuracy of 74.5 % and 76.4 %, and an AUC of 0.79 and 0.76, respectively.  Standard clinical features could predict the therapy response with an accuracy of 69.1 % and an AUC of 0.6.  Long-term survival analyses indicated that the patients predicted by the developed model as responders had a significantly better survival compared to the non-responders.  Similar findings were observed for the 2 response cohorts identified at post-treatment based on standard clinical and pathological criteria.  The authors concluded that the findings of this study demonstrated the potential of QUS multi-parametric imaging integrated with un-supervised learning methods in identifying distinct intra-tumor regions in breast cancer to characterize its responsiveness to chemotherapy before the commencement of treatment.  These researchers stated that the results obtained in this study are encouraging; however, studies on larger patient populations are needed to further examine the effectiveness and robustness of the proposed methodology in clinic.  Availability of data from larger patient cohorts will also allow stratified analyses across breast cancer sub-types to examine the performance of the developed models for different molecular and histological sub-types.

An UpToDate review on “Diagnostic evaluation of suspected breast cancer” (Esserman and Joe, 2021) does not mention non-elastographic quantitative ultrasound tissue characterization as a management option. 

Furthermore, National Comprehensive Cancer Network’s clinical practice guideline on “Breast cancer” (Version 2.2022) does not mention non-elastographic quantitative ultrasound tissue characterization as a management option.

Opto-Acoustic Imaging

Zalev et al (2019) noted that opto-acoustic imaging entails the use of light to produce sound waves for visualizing blood in biological tissue.  By means of multiple optical wavelengths, diagnostic images of blood oxygen saturation as well as total hemoglobin (Hb) are generated using endogenous optical contrast, without injection of any external contrast agent and without using any ionizing radiation.  The technology has been employed in recent clinical studies for diagnosis of breast cancer to aid in distinguishing benign from malignant lesions, potentially reducing the need for biopsy via improved diagnostic imaging accuracy.  To enable this application, techniques for mapping oxygen saturation differences within tissue are necessary.  Using biologically relevant opto-acoustic phantoms, these researchers analyzed the ability of an opto-acoustic imaging system to display colorized parametric maps that were generated using a statistical mapping approach.  To mimic breast tissue, a material with closely matching properties for optical absorption, optical scattering, acoustic attenuation, and speed of sound was used.  The phantoms include 2 vessels filled with whole blood at oxygen saturation levels determined using a sensor-based approach.  A flow system with gas-mixer and membrane oxygenator adjusted the oxygen saturation of each vessel independently.  Datasets were collected with an investigational Imagio breast imaging system.  These researchers examined the ability to distinguish vessels as the oxygen saturation level and imaging depth were varied.  At depth of 15 mm and hematocrit (Hct) of 42 %, a sufficient level of contrast to distinguish between two 1.6-mm diameter vessels was measured for an oxygen saturation difference of approximately 4.6 %.  Furthermore, an oxygenated vessel was visible at a depth of 48 mm using an optical wavelength of 1,064 nm, and a deoxygenated vessel was visible to a depth of 42 mm with 757 nm.  The authors concluded that by analyzing the effects of varying vessel depth and Hct level, it was determined that a reduction of contrast-to-background ratio (CBR) increased the blood oxygen saturation (sO2) difference needed to distinguish vessels.  In addition, by analyzing the color mapping approach, these investigators examined how the spatial distributions of relative sO2 and total Hb concentration (ctHb) may potentially be used during clinical evaluation to visualize differences between benign and malignant lesions for breast cancer diagnosis.  Moreover, these researchers stated that future investigations entail carrying out further device characterization as well as developing techniques to improve visualization of diagnostic imaging features.

Nyayapathi and Xia (2019) stated that breast cancer is one of the leading causes for cancer-related deaths in women, and early detection is critical in improving survival rates.  Currently, X-ray mammogram is the only modality for mass screening of asymptomatic women; however, it has decreased sensitivity in radiographically dense breasts, which is also associated with a higher risk for breast cancer.  Photo-acoustic (PA) imaging is an emerging modality that enables deep tissue imaging of optical contrast at ultrasonically defined spatial resolution, which is much higher than that can be achieved in purely optical imaging modalities.  Because of high optical absorption from Hb molecules, PA imaging could map out hemo distribution and dynamics in breast tissue and identify malignant lesions based on tumor associated angiogenesis and hypoxia.  These investigators examined various PA breast imaging systems proposed over the past few years and summarized the PA features of breast cancer identified in these systems.  The authors concluded that PA imaging has 2 major challenges.  The 1st challenge is optical attenuation in tissue.  Current results indicated that using multiple light illumination angles, along with optimizing imaging geometry could improve light penetration.  In addition to that, engineering PA-based contrast agents could be another solution to facilitate deep tissue imaging.  Second, although multi-spectral PA imaging allowed the functional quantification of breast tissue, the complexity of light absorption and scattering in tissue made the quantification inaccurate.  So far, only the relative values of oxygenated Hb (HbO2), Hb, and sO2 have been reported; thus, these functional metrics need to be thoroughly examined and standardized in order to pave the way for clinical translation.  These investigators stated that along with existing results and continuous improvement in system designs and image reconstruction techniques, PA imaging holds great promise for clinical translation in the field of breast cancer screening and diagnosis.


References

The above policy is based on the following references:

  1. Allen WM, Chin L, Wijesinghe P, et al. Wide-field optical coherence micro-elastography for intraoperative assessment of human breast cancer margins. Biomed Opt Express. 2016;7(10):4139-4153.
  2. Alveryd A, Andersson I, Aspegren K, et al. Lightscanning versus mammography for the detection of breast cancer in screening and clinical practice. A Swedish multicenter study. Cancer. 1990;65(8):1671-1677.
  3. American Cancer Society [website]. Mammograms and other breast imaging procedures. Atlanta, GA: American Cancer Society; revised September 30, 2009. Available at: http://www.cancer.org/docroot/CRI/content/CRI_2_6X_
    Mammography_and_other_Breast_Imaging_Procedures_5.asp. Accessed on March 24, 2010.
  4. Athanasiou A, Vanel D, Fournier L, Balleyguier C. Optical mammography: A new technique for visualizing breast lesions in women presenting non palpable BIRADS 4-5 imaging findings: Preliminary results with radiologic-pathologic correlation. Cancer Imaging. 2007;7:34-40.
  5. Au FW, Ghai S, Lu FI, Lu H. Clinical value of shear wave elastography added to targeted ultrasound (second-look ultrasound) in the evaluation of breast lesions suspicious of malignancy detected on magnetic resonance imaging. J Ultrasound Med. 2019;38(9):2395-2406.
  6. Balleyguier C, Lakhdar AB, Dunant A, et al. Value of whole breast magnetic resonance elastography added to MRI for lesion characterization. NMR Biomed. 2018;31(1).
  7. Bassett LW, Hendrick RE, Bassford TL, et al. Quality determinants of mammography. Clinical Practice Guideline No. 13. AHCPR Publication No. 95-0632. Rockville, MD: Agency for Health Care Policy and Research; October 1994.
  8. BC Cancer Agency. Breast - diagnosis. Types of cancer and sequelae. Patient/Public Information. Vancouver, BC; BC Cancer Agency; revised November 2001.
  9. Blackmore KM, Knight JA, Jong R, Lilge L. Assessing breast tissue density by transillumination breast spectroscopy (TIBS): An intermediate indicator of cancer risk. Br J Radiol. 2007;80(955):545-556.
  10. Blackmore KM, Knight JA, Lilge L. Association between transillumination breast spectroscopy and quantitative mammographic features of the breast. Cancer Epidemiol Biomarkers Prev. 2008;17(5):1043-1050.
  11. Blank MA, Antaki JF. Breast lesion elastography region of interest selection and quantitative heterogeneity: A systematic review and meta-analysis. Ultrasound Med Biol. 2017;43(2):387-397.
  12. Braddick MR. Audit of a breast cancer screening programme using clinical examination and lightscanning. Health Bull (Edinb). 1991;49(6):299-303.
  13. Byra M, Nowicki A, Wroblewska-Piotrzkowska H, Dobruch-Sobczak K. Classification of breast lesions using segmented quantitative ultrasound maps of homodyned K distribution parameters.  Med Phys. 2016;43(10):5561.
  14. Center for Devices and Radiological Health. Summary minutes of the 45th meeting of the obstetrics and gynecology devices panel (draft). Rockville, Maryland: US Food and Drug Administration, Center for Devices and Radiological Health, 1991.
  15. Centers for Disease Control. Inappropriate use of transillumination for breast cancer screening -- Wisconsin, 1990. MMWR Morb Mortal Wkly Rep. 1991;40(18):293-296.
  16. Centers for Medicare & Medicaid Services (CMS).  National Coverage Determination (NCD) for Transillumination Light Scanning or Diaphanography (30.9). Baltimore, MD: CMS; undated.
  17. Cespedes I, Ophir J, Ponnekanti H, Maklad N. Elastography: Elasticity imaging using ultrasound with application to muscle and breast in vivo. Ultrason Imaging. 1993;15:73-88.
  18. Chamming's F, Chopier J, Mathelin C, Chereau E. Explorations of breast microcalcifications: Guidelines. J Gynecol Obstet Biol Reprod (Paris). 2015;44(10):960-969.
  19. Chauveau N, Hamzaoui L, Rochaix P, et al. Ex vivo discrimination between normal and pathological tissues in human breast surgical biopsies using bioimpedance spectroscopy. Ann N Y Acad Sci. 1999;873:42-50.
  20. Chen L, He J, Liu G, et al. Diagnostic performances of shear-wave elastography for identification of malignant breast lesions: A meta-analysis. Jpn J Radiol. 2014;32(10):592-599.
  21. Chen W, Fang L-X, Chen H-L, Zheng J-H. Accuracy of ultrasound elastography for predicting breast cancer response to neoadjuvant chemotherapy: A systematic review and meta-analysis. World J Clin Cases. 2022;10(11):3436-3448.
  22. Cherepenin V, Karpov A, Korjenevsky A, et al. A 3D electrical impedance tomography (EIT) system for breast cancer detection. Physiol Meas. 2001;22(1):9-18.
  23. Cherepenin VA, Karpov AY, Korjenevsky AV, et al. Three-dimensional EIT imaging of breast tissues: System design and clinical testing. IEEE Trans Med Imaging. 2002;21(6):662-667.
  24. Cho N, Moon WK, Park JS, et al. Nonpalpable breast masses: Evaluation by US elastography. Korean J Radiol. 2008;9(2):111-118.
  25. Chung J, Lee WK, Cha ES, et al. Shear-wave elastography for the differential diagnosis of breast papillary lesions. PLoS One. 2016;11(11):e0167118.
  26. Collins LC, Laronga C, Wong JS. Microinvasive breast carcinoma. UpToDate [online serial]. Waltham, MA: UpToDate; reviewed February 2018.
  27. Crombe A, Hurtevent-Labrot G, Asad-Syed M, et al. Shear-wave elastography quantitative assessment of the male breast: Added value to distinguish benign and malignant palpable masses. Br J Radiol. 2018;91(1082):20170676.
  28. de Faria Castro Fleury E, Gianini AC, Ayres V, et al. Application of breast ultrasound elastography to differentiate intracapsular collection from silicone-induced granuloma of breast implant capsule complementarily to contrast-enhanced breast magnetic resonance imaging. Breast Cancer (Auckl). 2017;11:1178223417737994.
  29. Di Meo S, Pasotti L, Lashkevich E, et al. Combining millimeter-wave imaging, ultrasound and elastography in a new multimodal approach for breast cancer detection: Initial experimental results. Annu Int Conf IEEE Eng Med Biol Soc. 2020;2020:1807-1810.
  30. Dong F, Wu H, Zhang L, et al. Diagnostic performance of multimodal sound touch elastography for differentiating benign and malignant breast masses. J Ultrasound Med. 2019;38(8):2181-2190.
  31. Esserman LJ, Joe BN. Diagnostic evaluation of suspected breast cancer. UpToDate [online serial]. Waltham, MA: UpToDate; reviewed December 2021.
  32. Filip I, Ferre R, AlSharif S, et al. Can strain elastography improve the characteri.zation of breast lesions identified during second-look MRI-directed sonographic examination? J Clin Ultrasound. 2019;47(4):187-194.
  33. Fleury Ede F, Fleury JC, Piato S, Roveda D Jr. New elastographic classification of breast lesions during and after compression. Diagn Interv Radiol. 2009;15(2):96-103.
  34. Forte AJ, Huayllani MT, Boczar D, et al. The basics of ultrasound elastography for diagnosis, assessment, and staging breast cancer-related lymphedema: A systematic review of the literature. Gland Surg. 2020;9(2):589-595.
  35. Fowlkes JB, Emelianov SY, Pipe JG, et al. Magnetic-resonance imaging techniques for detection of elasticity variation. Med Phys. 1995;22(11 Pt 1):1771-1778.
  36. Freer PE, Slanetz PJ. Breast density and screening for breast cancer. UpToDate [online serial]. Waltham, MA: UpToDate; reviewed February 2018.
  37. Gandjbakhche AH, Nossal R, Bonner RF. Resolution limits for optical transillumination of abnormalities deeply embedded in tissues. Med Phys. 1994;21(2):185-191.
  38. Garra BS, Cespedes EI, Ophir J, et al. Elastography of breast lesions: Initial clinical results. Radiology. 1997;202:79-86.
  39. Garra BS. Imaging and estimation of tissue elasticity by ultrasound. Ultrasound Q. 2007;23(4):255-268.
  40. Giuseppetti GM, Martegani A, Di Cioccio B, Baldassarre S. Elastosonography in the diagnosis of the nodular breast lesions: Preliminary report. Radiol Med. 2005;110(1-2):69-76.
  41. Glickman YA, Filo O, Nachaliel U, et al. Novel EIS postprocessing algorithm for breast cancer diagnosis. IEEE Trans Med Imaging. 2002;21(6):710-712.
  42. Gonzalez J, Decerce J, Erickson SJ, et al. Hand-held optical imager (Gen-2): Improved instrumentation and target detectability. J Biomed Opt. 2012;17(8):081402-1.
  43. Gordenne W, Bauduin E. Diagnostic accuracy of new imaging techniques in breast diseases. J Belge Radiol. 1989;72(1):35-38.
  44. Gubarkova EV, Kiseleva EB, Sirotkina MA, et al. Diagnostic accuracy of cross-polarization OCT and OCT-elastography for differentiation of breast cancer subtypes: Comparative study. Diagnostics (Basel). 2020;10(12):994.
  45. Guo G, Feng J, Jin C, et al. A novel nomogram based on imaging biomarkers of shear wave elastography, angio planewave ultrasensitive imaging, and conventional ultrasound for preoperative prediction of malignancy in patients with breast lesions. Diagnostics (Basel). 2023;13(3):540.
  46. Hayes JC. Electrical impedance images for breast gain FDA approval. Diagn Imaging (San Franc). 1999;21(6):19-20.
  47. Heywang-Kobrunner SH. Nonmammographic breast imaging techniques. Curr Opin Radiol. 1992;4(5):146-154.
  48. Huang X-W, Huang Q-X, Huang H, et al. Diagnostic performance of quantitative and qualitative elastography for axillary lymph node metastasis in breast cancer: A systematic review and meta-analysis. Front Oncol. 2020;10:552177.
  49. Ianculescu V, Ciolovan LM, Dunant A, et al. Added value of Virtual Touch IQ shear wave elastography in the ultrasound assessment of breast lesions. Eur J Radiol. 2014;83(5):773-777.
  50. Imaging Diagnostic Systems, Inc [website]. Computed tomography laser mammography (CTLM). Fort Lauderdale, FL: Imaging Diagnostic Systems, Inc., 2010. Available at: http://www.imds.com/. Accessed on March 24, 2010.
  51. Itoh A, Ueno E, Tohno E, et al. Breast disease: Clinical application of US elastography for diagnosis. Radiology. 2006;239: 341-350.
  52. Jackson VP, Hendrick RE, Feig SA, et al. Imaging of the radiographically dense breast. Radiology. 1993;188(2):297-301.
  53. Jales RM, Dória MT, Serra KP, et al. Power Doppler ultrasonography and shear wave elastography as complementary imaging methods for suspected local breast cancer recurrence. J Ultrasound Med. 2018;37(6):1493-1501.
  54. Jarlman O, Andersson I, Balldin G, et al. Diagnostic accuracy of lightscanning and mammography in women with dense breasts. Acta Radiol. 1992;33(1):69-71.
  55. Jarlman O, Balldin G, Andersson I, et al. Relation between lightscanning and the histologic and mammographic appearance of malignant breast tumors. Acta Radiol. 1992;33(1):63-68.
  56. Jossinet J, Schmitt M. A review of parameters for the bioelectrical characterization of breast tissue. Ann N Y Acad Sci. 1999;873:30-41.
  57. Jung E, Hosl V, von Fraunberg S, et al. Ultrasound elastography for the detection of capsular fibrosis in breast implants: First results. Clin Hemorheol Microcirc. 2021;77(3):247-257.
  58. Kennedy KM, Chin L, Wijesinghe P, et al. Investigation of optical coherence micro-elastography as a method to visualize micro-architecture in human axillary lymph nodes. BMC Cancer. 2016;16(1):874.
  59. Kerne TE, Hartov A, Soho SK, et al. Imaging the breast with EIS: an initial study of exam consistency. Physiol Meas. 2002;23(1):221-236.
  60. Kerner TE, Paulsen KD, Hartov A, et al. Electrical impedance spectroscopy of the breast: Clinical imaging results in 26 subjects. IEEE Trans Med Imaging. 2002;21(6):638-645.
  61. Kim GR, Choi JS, Han BK, et al. Combination of shear-wave elastography and color Doppler: Feasible method to avoid unnecessary breast excision of fibroepithelial lesions diagnosed by core needle biopsy. PLoS One. 2017;12(5):e0175380.
  62. Kostli KP, Frenz M, Weber HP, et al. Optoacoustic tomography: Time-gated measurement of pressure distributions and image reconstruction. Appl Opt. 2001;40(22):3800-3809.
  63. Kumm TR, Szabunio MM. Elastography for the characterization of breast lesions: Initial clinical experience. Cancer Control. 2010;17(3):156-161.
  64. Kyle JR, Kyle MD, Raghavan R, et al. Analysis of light scattering from human breast tissue using a custom dual-optical scanning near-field optical microscope. J Biophotonics. 2011;4(3):193-205.
  65. Landoni V, Francione V, Marzi S, et al. Quantitative analysis of elastography images in the detection of breast cancer. Eur J Radiol. 2012;81(7):1527-1531.
  66. Li DD, Guo LH, Xu HX, et al. Acoustic radiation force impulse elastography for differentiation of malignant and benign breast lesions: A meta-analysis. Int J Clin Exp Med. 2015;8(4):4753-4761.
  67. Liu B, Zheng Y, Huang G, et al. Breast lesions: Quantitative diagnosis using ultrasound shear wave elastography -- A systematic review and meta-analysis. Ultrasound Med Biol. 2016;42(4):835-847.
  68. Liufu V, Hiller J. Breast cancer diagnosis using ultrasound elasticity imaging. Prioritising Summary Update. Canberra, ACT: Adelaide Health Technology Assessment; November 2008.
  69. Lorenzen J, Sinkus R, Adam G. [Elastography: Quantitative imaging modality of the elastic tissue properties] Rofo. 2003;175(5):623-630.
  70. Lorenzen J, Sinkus R, Lorenzen M, et al. Elastography of the breast: Preliminary clinical results. Rofo. 2002;174(7):830-834.
  71. Ma Y, Zhang S, Zang L, et al. Combination of shear wave elastography and Ki-67 index as a novel predictive modality for the pathological response to neoadjuvant chemotherapy in patients with invasive breast cancer. Eur J Cancer. 2016 69:86-101.
  72. Malich A, Boehm T, Facius M, et al. Differentiation of mammographically suspicious lesions: Evaluation of breast ultrasound, MRI mammography and electrical impedance scanning as adjunctive technologies in breast cancer detection. Clin Radiol. 2001;56(4):278-283.
  73. Malich A, Bohm T, Facius M, et al. Additional value of electrical impedance scanning: Experience of 240 histologically-proven breast lesions. Eur J Cancer. 2001;37(18):2324-2330.
  74. Malich A, Fritsch T, Anderson R, et al. Electrical impedance scanning for classifying suspicious breast lesions: First results. Eur Radiol. 2000;10(10):1555-1561.
  75. Malich A, Fuchsjager M. Electrical impedance scanning in classifying suspicious breast lesions. Invest Radiol. 2003;38(5):302-304.
  76. Manduca A, Oliphant TE, Dresner MA, et al. Magnetic resonance elastography: Non-invasive mapping of tissue elasticity. Med Image Anal. 2001;5(4):237-254.
  77. Martin G, Martin R, Brieva MJ, Santamaria L. Electrical impedance scanning in breast cancer imaging: Correlation with mammographic and histologic diagnosis. Eur Radiol. 2002;12(6):1471-1478.
  78. McKnight AL, Kugel JL, Rossman PJ, et al. MR elastography of breast cancer: Preliminary results. AJR. 2002;178(6):1411-1417.
  79. Moon WK, Choi JW, Cho N, et al. Computer-aided analysis of ultrasound elasticity images for classification of benign and malignant breast masses. AJR Am J Roentgenol. 2010;195(6):1460-1465.
  80. Mundy L, Liufu V, Braunack-Mayer A, et al. New and emerging technologies for breast cancer detection. National Horizon Scanning Emerging Technology Bulletin. Canberra, ACT: Adelaide Health Technology Assessment; February 2009.
  81. National Academy of Sciences, Institute of Medicine, National Cancer Policy Board, Committee on the Early Detection of Breast Cancer. Mammography and Beyond: Developing Technologies for the Early Detection of Breast Cancer. Washington, DC: National Academy Press; 2001.
  82. National Comprehensive Cancer Network (NCCN). Breast cancer. NCCN Clinical Practice Guidelines in Oncology, Version 2.2022. Plymouth Meeting, PA: NCCN; 2022.
  83. No authors listed. Reassessment of transillumination light scanning for the diagnosis of breast cancer. Health Technol Assess Rep. 1988;(2):1-7.
  84. Nyayapathi N, Xia J. Photoacoustic imaging of breast cancer: A mini review of system design and image features. J Biomed Opt. 2019;24(12):1-13.
  85. Ohmine Y, Morimoto T, Kinouchi Y, et al. Noninvasive measurement of the electrical bioimpedance of breast tumors. Anticancer Res. 2000;20(3B):1941-1946.
  86. Olgun DC, Korkmazer B, Kilic F, et al. Use of shear wave elastography to differentiate benign and malignant breast lesions. Diagn Interv Radiol. 2014;20(3):239-244.
  87. Ophir J, Céspedes I, Ponnekanti H, et al. Elastography: A quantitative method for imaging the elasticity of biological tissues. Ultrason Imaging. 1991;13:111–134.
  88. Osapoetra LO, Chan W, Tran W, et al. Comparison of methods for texture analysis of QUS parametric images in the characterization of breast lesions. PLoS One. 2020;15(12):e0244965.
  89. Osapoetra LO, Sannachi L, Quiaoit K, et al. A priori prediction of response in multicentre locally advanced breast cancer (LABC) patients using quantitative ultrasound and derivative texture methods. Oncotarget. 2021;12(2):81-94.
  90. Parekh A, Das D, Das S, et al. Bioimpedimetric analysis in conjunction with growth dynamics to differentiate aggressiveness of cancer cells. Sci Rep. 2018;8(1):783.
  91. Park S-Y, Kang BJ. Combination of shear-wave elastography with ultrasonography for detection of breast cancer and reduction of unnecessary biopsies: A systematic review and meta-analysis. Ultrasonography. 2021;40(3):318-332.
  92. Perlet C, Kessler M, Lenington S, et al. Electrical impedance measurement of the breast: Effect of hormonal changes associated with the menstrual cycle. Eur Radiol. 2000;10(10):1550-1554.
  93. Pillai A, Voruganti T, Barr R, Langdon J. Diagnostic accuracy of shear-wave elastography for breast lesion characterization in women: A systematic review and meta-analysis. J Am Coll Radiol. 2022;19(5):625-634.
  94. Poellinger A, Martin JC, Ponder SL, et al. Near-infrared laser computed tomography of the breast first clinical experience. Acad Radiol. 2008;15(12):1545-1553.
  95. Radai MM, Abboud S, Rosenfeld M. Evaluation of impedance technique for detecting breast carcinoma using a 2-D numerical model of the torso. Ann N Y Acad Sci. 1999;873:360-369.
  96. Sabel M, Aichinger H. Recent developments in breast imaging. Phys Med Biol. 1996;41(3):315-368.
  97. Sadigh G, Carlos RC, Neal CH, Dwamena BA. Accuracy of quantitative ultrasound elastography for differentiation of malignant and benign breast abnormalities: A meta-analysis. Breast Cancer Res Treat. 2012;260(2):428-436.
  98. Seo M, Choi HY, Sohn YM, et al. Shear wave elastography for the diagnosis of small (≤2 cm) breast lesions: Added value and factors associated with false results. Br J Radiol. 2019;92(1097):20180341.
  99. Shahzad R, Fatima I, Anjum T, Shahid A. Diagnostic value of strain elastography and shear wave elastography in differentiating benign and malignant breast lesions. Ann Saudi Med. 2022;42(5):319-326.
  100. Sinkus R, Tanter M, Xydeas T, et al. Viscoelastic shear properties of in vivo breast lesions measured by MR elastography. Magn Reson Imaging. 2005;23(2):159-165.
  101. Souchon R, Soualmi L, Bertrand M, et al. Ultrasonic elastography using sector scan imaging and a radial compression. Ultrasonics. 2002;40(1-8):867-871.
  102. Stojadinovic A, Nissan A, Gallimidi Z, et al. Electrical impedance scanning for the early detection of breast cancer in young women: Preliminary results of a multicenter prospective clinical trial. J Clin Oncol. 2005;23(12):2703-2715.
  103. Stojadinovic A, Nissan A, Shriver CD, et al. Electrical impedance scanning as a new breast cancer risk stratification tool for young women. J Surg Oncol. 2008;97(2):112-120.
  104. Taleghamar H, Moghadas-Dastjerdi H, Czarnota GJ, Sadeghi-Naini A. Characterizing intra-tumor regions on quantitative ultrasound parametric images to predict breast cancer response to chemotherapy at pre-treatment. Sci Rep. 2021;11(1):14865.
  105. Tardivon A, El Khoury C, Thibault F, et al. [Elastography of the breast: A prospective study of 122 lesions]. J Radiol. 2007;88(5 Pt 1):657-662.
  106. Taulbee P. FDA panel rejects Mirabel's bioimpedance device for breast scanning. FDC Reports: The Gray Sheet. 2006;32(36):8.
  107. Thomas A, Fischer T, Frey H, et al. Real-time elastography--an advanced method of ultrasound: First results in 108 patients with breast lesions. Ultrasound Obstet Gynecol. 2006;28(3):335-340.
  108. Thomas A, Kümmel S, Fritzsche F, et al. Real-time sonoelastography performed in addition to B-mode ultrasound and mammography: Improved differentiation of breast lesions? Acad Radiol. 2006;13(12):1496-1504.
  109. Toprak N, Toktas O, Ince S, et al. Does ARFI elastography complement B-mode ultrasonography in the radiological diagnosis of idiopathic granulomatous mastitis and invasive ductal carcinoma? Acta Radiol. 2022;63(1):28-34.
  110. Tromberg BJ, Cerussi A, Shah N, et al. Imaging in breast cancer: Diffuse optics in breast cancer: Detecting tumors in pre-menopausal women and monitoring neoadjuvant chemotherapy. Breast Cancer Res. 2005;7(6):279-285.
  111. Tromberg BJ, Pogue BW, Paulsen KD, et al. Assessing the future of diffuse optical imaging technologies for breast cancer management. Med Phys. 2008;35(6):2443-2451.
  112. U.S. Food and Drug Administration (FDA) 510(k). Regen SonixTouch Ultrasound Scanner. Summary of Safety and Effectiveness. 510(k) No. K083095. Rockville, MD: FDA. October 31, 2008.
  113. U.S. Food and Drug Administration (FDA). FDA approves new breast imaging device. FDA Talk Paper. T99-18. Rockville, MD: FDA; April 19, 1999.
  114. Vreugdenburg TD, Willis CD, Mundy L, Hiller JE. A systematic review of elastography, electrical impedance scanning, and digital infrared thermography for breast cancer screening and diagnosis. Breast Cancer Res Treat. 2013;137(3):665-676.
  115. Wang F, Wang H. Diagnostic value of ultrasound elastography in triple negative breast cancer: A meta-analysis. Medicine (Baltimore). 2023;102(6):e32879.
  116. Wang T, Wang K, Yao Q, et al. Prospective study on combination of electrical impedance scanning and ultrasound in estimating risk of development of breast cancer in young women. Cancer Invest. 2010;28(3):295-303.
  117. Wang W, Tang M, McCormick M, Dong X. Preliminary results from an EIT breast imaging simulation system. Physiol Meas. 2001;22(1):39-48.
  118. Wang XL, Tao L, Zhou XL, et al. Initial experience of automated breast volume scanning (ABVS) and ultrasound elastography in predicting breast cancer subtypes and staging. Breast. 2016;30:130-135.
  119. Wersebe A, Siegmann K, Krainick U, et al. Diagnostic potential of targeted electrical impedance scanning in classifying suspicious breast lesions. Invest Radiol. 2002;37(2):65-72.
  120. Wu H, Liang W, Jiao Y, et al. A preliminary comparative study of Young's modulus versus shear modulus in the diagnosis of breast cancer. Ultrasound Q. 2019;35(1):88-92.
  121. Wu H, Wang C, An Q, et al. Comparing the accuracy of shear wave elastography and strain elastography in the diagnosis of breast tumors: A systematic review and meta-analysis. Medicine (Baltimore). 2022;101(44):e31526.
  122. Xue Y, Yao S, Li X, Zhang H. Value of shear wave elastography in discriminating malignant and benign breast lesions: A meta-analysis. Medicine (Baltimore). 2017;96(42):e7412.
  123. Xydeas T, Siegmann K, Sinkus R, et al. Magnetic resonance elastography of the breast: Correlation of signal intensity data with viscoelastic properties. Invest Radiol. 2005;40(7):412-420.
  124. Yang K, Ye X, Tian H, et al. Development and validation of a nomogram for discriminating between benign and malignant breast masses by conventional ultrasound and dual-mode elastography: A multicenter study. Quant Imaging Med Surg. 2023;13(2):865-877.
  125. Yi A, Cho N, Chang JM, et al. Sonoelastography for 1786 non-palpable breast masses: Diagnostic value in the decision to biopsy. Eur Radiol. 2012;22(5):1033-1040.
  126. Youk JH, Gweon HM, Son EJ. Shear-wave elastography in breast ultrasonography: The state of the art. Ultrasonography. 2017;36(4):300-309.
  127. Zalev J, Richards LM, Clingman BA, et al. Opto-acoustic imaging of relative blood oxygen saturation and total hemoglobin for breast cancer diagnosis. J Biomed Opt. 2019;24(12):1-16.
  128. Zhang XF, Liu XM, Bao XF, et al. [Application of real-time tissue elastography in diagnosis of breast cancer]. Zhejiang Da Xue Xue Bao Yi Xue Ban. 2006;35(4):444-447.
  129. Zou Y, Guo Z. A review of electrical impedance techniques for breast cancer detection. Med Eng Phys. 2003;25(2):79-90.